Bugbox
UM-16 R5.0
Page 38 of 48
Bugbox User Manual
Replacing the detox and catalyst sachets
A detox and a catalyst sachet are supplied with the Bugbox workstation. The detox sachet adsorbs
volatile organic compounds, improving the air quality within the workstation chamber. The catalyst
sachet contains the catalyst required for the hydrogen to react with any oxygen in the workstation, to
produce an anoxic environment. The detox sachet and the catalyst sachet need to be replaced
annually. See section 7.4.2 for ordering details. It is recommended to replace both sachets at the same
time. Note: Both sachets are replaced as part of an annual service.
To replace the detox sachet and the catalyst sachet;
Remove any samples to an alternative storage facility.
Switch off Bugbox and disconnect from the mains power supply.
Remove the Ezee Sleeves (see section 7.2.2 for more information).
Undo the glove port covers and place on the rear shelf.
Lift the floor tray at the front to expose the detox and catalyst sachets. Figure 27 shows the
location of the sachet holders;
Figure 27: Sachet holder location
1.
Floor tray.
2.
Sachet holder. The second sachet holder is located behind and to the
left of the first sachet holder.
3.
Floor tray locator.
Slide the sachets forwards to remove them from their holders.
Remove the sachets via the glove ports.
Lower the floor tray, ensuring that the floor tray leg has located in the floor tray locator (item
3 in Figure 27).
1
2
3