Chapter 7. Foundation Fieldbus Communications
7.5 Resetting Instrument Totalizers
To reset the instrument totalizers:
1.
Double click on the FLOW transducer block (in the tree under GEFlow-XMT; see Figure 48 on page 66).
2.
Select the
Others
tab and scroll down to the CLEAR_TOTALIZERS listing.
3.
Select
Clear
from the drop down list box (see Figure 51).
4.
After the totals have been reset, select
Normal
from the drop down list box to resume total accumulation.
Figure 51: Clear Totalizers Drop Down List