6.5.2 The opt>setup menu
This menu is used to restore corrupted software or to
download new software to the XDP. The following items
are required:
• A hardware connection between the XDP serial port
and the serial port on a personal computer
• The new main code and slot code software files, as
provided by Panametrics
• An ASCII data transfer program, such as windows
Hyperterminal When you have all of the above items,
proceed as follows:
6.5.2a Loading the new main code
The first step in the process is to download the new
main code to the XDP. To do this, navigate to the
opt>setup
menu on the XDP and proceed as follows:
1. From the PC, open the data transfer program and
configure the serial port as follows: baud rate = 19200,
data bits = 8, stop bits = 1, parity = none, and hardware
flow control = none.
2. From the XDP, at the
setup
menu, press [ENTER].
3. Use the XDP [
] and [
] keys to highlight the
load main
option and press [ENTER].
4. From the PC, when asked, press the
<Y>
key
to proceed.
5. Select
<transfer>
and then
<send file>
.
6. Browse to the
main code *.cod
file provided by
Panametrics, and select
<Xmodem>
for the transfer
protocol.
7. Press
<send>
.
After the new main code has been downloaded, the
XDP will reboot. Re-enter the factory-level password and
proceed to the next section.
6.5.2b Erasing the old main code and slot code
The next step in the process is to erase the old software
from the XDP. To do this, navigate to the
opt>setup
menu
on the XDP and proceed as follows:
1. At the
setup
menu, press [ENTER].
2. Use the [
] and [
] keys to highlight the
erase main
option and press [ENTER]. All the current customer
configuration data will be erased and the XDP will reboot.
3. During the reboot process,
scanning hardware
will
appear on the XDP display. When it does, use the
[
] and [
] keys to select
skip slot download
and then
select
yes
.
4. Re-enter the factory-level password when prompted,
and the XDP will reboot again. When the reboot is
complete, re-enter the factory-level password and
navigate to the
opt>setup
menu
5. Use the [
] and [
] keys to highlight the
erase slot
option and press [ENTER]. The old slot code will be
erased and the XDP will reboot.
Re-enter the factory-level password and proceed to the
next section.
6.5.2c Loading the new slot code
The final step in the process is to load the new slot code.
To do this, navigate to the opt>setup menu on the XDP and
proceed as follows:
1. At the
setup
menu, press [ENTER].
2. Use the XDP [
] and [
] keys to highlight the
load slot
option and press [ENTER].
3. From the PC, when asked, press the
<Y>
key
to proceed.
4. Select <transfer> and then
<Send File>
.
5. Browse to the
slot code *.cod
file provided by
Panametrics, and select
<Xmodem>
for the transfer
protocol.
6. Press
<send>
.
After the new slot code has been downloaded, the
XDP will reboot. The XDP software upgrade has now been
completed, but all programming options will now be in
the factory default state. You will need to reprogram the
unit with your preferred settings.
Содержание Panametrics XDP
Страница 1: ...Gas analysis XDP User s manual 910 204 Rev E...
Страница 2: ...ii...
Страница 4: ...no content intended for this page proceed to next page iv...
Страница 8: ...viii no content intended for this page...
Страница 12: ...no content intended for this page proceed to next page 2...
Страница 28: ...no content intended for this page proceed to next page 18...
Страница 36: ...26 no content intended for this page proceed to next page...
Страница 42: ...32 no content intended for this page proceed to next page...
Страница 58: ...48 no content intended for this page proceed to next page...