17
3.5 Accessing the user program
To configure the XDP for a specific application, it is
necessary to leave measurement mode and enter the
user program
as follows (see menu maps in
and
):
While in measurement
mode, press the [ENTER],
[ENTER], [ESC], [ENTER], and
[ESC] keys in sequence.
Enter the operator-level password (
2719
) at the next
screen to access the menus available to the end user.
Note that a different password is required to access the
factory-level menus.
Use [
] and [
] to
position the cursor under
the desired character.
Then, use [
] and [
] to
increment the value. Press
[ENTER] when done.
This screen shows the
highest level of the
user
program
, which is referred
to as the
main menu
.
3.6 Exiting the user program
To leave the
user program
, proceed as follows from the
main menu:
Press the [ESC] key.
You are now back in
normal measurement
mode.
Notice that the padlock is no longer visible at the upper
right corner of the display window. Once the
user program
has been accessed with a valid password, the menus at
that programming level remain unlocked when you return
to measurement mode. With the menus unlocked, you
may reenter the user program by simply pressing the
[ESC] key only. To re-lock the menus, see the instructions
in
.
NOTE:
If the unit is powered down, the
user program
will be
locked when the unit is powered up again.
To program a specific feature of the XDP, proceed to the
appropriate section of
.
3.7 Error codes
As an aid in monitoring the status of your system, the XDP may display a variety of blinking error codes on the LCD.
These error codes are listed in
Table 1
below.
Table 1: Error code descriptions
Code
Meaning
Code
Meaning
LL
Low low alarm has been tripped
(low alarm normally also tripped)
Input over range
Analog input is above the
programmed range
L
Low alarm has been tripped
Zero drift/cal
Zero drift/cal is over
programmed limit
H
High alarm has been tripped
Zero drift/total
Zero drift/total is over
programmed limit
HH
High high alarm has been tripped
(high alarm normally also tripped)
Span drift/cal
Span drift/cal is over
programmed limit
CV
Auto verification alarm has
been tripped
Span drift/total
Span drift/total is over
programmed limit
Input under range
Analog input is below
the programmed range
Содержание Panametrics XDP
Страница 1: ...Gas analysis XDP User s manual 910 204 Rev E...
Страница 2: ...ii...
Страница 4: ...no content intended for this page proceed to next page iv...
Страница 8: ...viii no content intended for this page...
Страница 12: ...no content intended for this page proceed to next page 2...
Страница 28: ...no content intended for this page proceed to next page 18...
Страница 36: ...26 no content intended for this page proceed to next page...
Страница 42: ...32 no content intended for this page proceed to next page...
Страница 58: ...48 no content intended for this page proceed to next page...