oxy.IQ User’s Manual
19
Chapter 6. Specifications
Chapter 6. Specifications
6.1
Intrinsically Safe (IS) Installation
Intrinsically safe installations require an MTL7706 Zener barrier, one IS cable, and one non-IS cable.
6.1.1
Power requirements
24 to 28 VDC at 50 mA
6.1.2
Cable
P/N 704-1318-02 (2 m length) or p/n 704-1317-10 (10 m length) blue jacketed, twisted-pair with connector, 26 AWG
conductors, with connector
6.1.3
Output
Total load must equal 250 ±5% when using a Zener barrier
6.2 Non-incendive (div 2) and general purpose installation
No Zener barrier or galvanic isolator is used.
6.2.1
Power requirements
9 to 28 VDC, loop-powered, 0.7 W max
6.2.2 Cable
P/N 704-1317-02 (2 m length) or p/n 704-1317-10 (10 m length) black jacketed, twisted-pair with connector, 26 AWG
conductors, with connector
6.3 All Installations
6.3.1
Process Wetted Materials
SS process unit: 316 stainless steel, Viton® o-ring, gold-plated sensor electrical contacts, and glass
6.3.2 User-Selectable Measurement Ranges
•
PPM sensors:
- 0 to 10 ppm
v
O
2
(OX-1 or OX-2 only)
- 0 to 20 ppm
v
O
2
(OX-1 or OX-2 only)
- 0 to 50 ppm
v
O
2
(OX-1 or OX-2 only)
- 0 to 100 ppm
v
O
2
- 0 to 200 ppm
v
O
2
- 0 to 500 ppm
v
O
2
- 0 to 1000 ppm
v
O
2
- 0 to 2000 ppm
v
O
2
- 0 to 5000 ppm
v
O
2
- 0 to 10,000 ppm
v
O
2
•
Percent sensors:
- 0% to 1% O
2
- 0% to 2% O
2
- 0% to 5% O
2
- 0% to 10% O
2
- 0% to 25% O
2
- 0% to 50% O
2
6.3.3 Accuracy
•
±1% of range at calibration point
•
±2% of range at the calibration point for the 0 to 10 ppm
v
O
2
range (OX-1, 2)
Содержание Panametrics oxy.IQ
Страница 1: ...panametrics com BH024C11 EN G oxy IQ Panametrics Oxygen Transmitter User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 12: ...Chapter 1 Features and Capabilities 4 oxy IQ User s Manual no content intended for this page...
Страница 26: ...Chapter 5 The Service Menu 18 oxy IQ User s Manual no content intended for this page...
Страница 30: ...Chapter 6 Specifications 22 oxy IQ User s Manual no content intended for this page...
Страница 38: ...Appendix A Outline and Installation Drawings 30 oxy IQ User s Manual no content intended for this page...
Страница 42: ...Appendix C Order String 34 oxy IQ User s Manual no content intended for this page...
Страница 50: ...Appendix E Cell Models 42 oxy IQ User s Manual no content intended for this page...
Страница 54: ...Warranty 48 oxy IQ User s Manual no content intended for this page...
Страница 55: ......