background image

2.3  Advanced setup

The following sections describe the procedures for completing the configuration of your 

HygroPro

 transmitter.

2.3.1  Setting up the pressure/temperature 

displays

The following steps set the displayed pressure and 

temperature values to live (changing with the current 

measurements) or constant (remaining the same, 

regardless of the current measurements). If constant is 

selected, the desired numerical value must be set.

To set the pressure display, 

enter your password 

and use the arrow keys 

to select 

const pressure

Press 

enter

.

Use the arrow keys to 

select either 

live pressure

 

or 

constant pressure

 and 

press 

enter

.

If you selected constant 

pressure, use the arrow 

keys to select press... and 

press 

enter

.

Use the arrow keys to 

enter

 the desired pressure 

value and press enter. The 

screen will be updated to 

show the new pressure 

value.

To set the temperature 

display, enter your 

password and use the 

arrow keys to select const 

temperature. Press 

enter

.

Use the same procedure 

to set the temperature 

mode, and if 

constant 

temperature

 is selected, 

to enter the constant 

temperature value.

2.3.2  Entering sensor calibration data

Enter your password and 

use the arrow keys to scroll 

to 

calibrate

. Press 

enter

.

Select 

Hygro curve

 and 

press 

enter

.

Select 

row 1 

and press 

enter

.

Select the desired units 

and press 

enter

.

Use the arrow keys to enter 

a value and press 

enter

.

Repeat the previous steps until all of your 

Hygro Curve

 data 

points have been entered.

To save the data you 

have entered, scroll to the 

bottom of the Hygro Curve 

options list to the SAVE 

HYGRO tABLE? section. 

Then, select save and 

press 

enter

.

IMPORTANT

Any data not saved at the above prompt will be lost.

14

Содержание Panametrics HygroPro

Страница 1: ...HygroPro User s manual 916 099 Rev D...

Страница 2: ...ii...

Страница 3: ...HygroPro Moisture transmitter User s manual 916 099 Rev D November 2012 iii...

Страница 4: ...no content intended for this page proceed to next page iv...

Страница 5: ...1 2 2 3 Selecting measurement parameters 11 2 2 4 Setting up the display 12 2 2 5 Setting up the analog output 12 2 2 6 Entering the node ID 13 2 2 7 Setting up the RS485 output 13 2 3 Advanced setup...

Страница 6: ...the cleaned probe 18 Chapter 4 Specifications 19 4 1 General 19 4 2 Electrical 19 4 3 Mechanical 19 4 4 Moisture sensor 19 4 5 Built in temperature sensor 20 4 6 Built in pressure sensor 20 4 7 Certi...

Страница 7: ...t the following link https www bakerhughesds com sales terms conditions Typographical conventions Note These paragraphs provide information that provides a deeper understanding of the situation but is...

Страница 8: ...ke sure that unauthorized personnel cannot gain access to the operation of the equipment Environmental compliance RoHS The HygroPro moisture transmitter fully complies with RoHS regulations Directive...

Страница 9: ...x button keypad and it is housed in an IP67 Type 4X enclosure The HygroPro transmitter uses an aluminum oxide moisture sensor and it includes a temperature thermistor and a pressure transducer on a co...

Страница 10: ...he sample flow Before constructing a sample system consult a Panametrics application engineer and observe the following guidelines The sample system should be very simple and should contain as few com...

Страница 11: ...age during operation 2 Using the integral 3 4 16 straight male thread screw the probe end of the transmitter into the process or sample system fitting Make sure not to damage the threads Note A 3 4 16...

Страница 12: ...receive and display the 4 20 mA measurement output from the HygroPro proceed to next page an external power supply to provide power to the HygroPro plus a personal computer PC running Panametrics Pan...

Страница 13: ...ground 1 Connect an RS485 RS232 converter customer supplied to an available serial port on the PC 2 Push the female connector end of the factory supplied cable into the mating male connector on the t...

Страница 14: ...da for the following hazardous classified locations and protection methods Intrinsically safe class I II III division 1 group ABCDEFG T4 T Ambient 40 C to 60 C Enclosure IP 67 3 Galvanically isolated...

Страница 15: ...C22 2 No 61010 1 Degrees of Protection Provided by Enclosures IP Code ANSI IEC 60529 Degrees of Protection Provided by Enclosures IP Code CSA C22 2 No 60529 1 4 2b Compliance requirements The HygroPr...

Страница 16: ...e end of the HygroPro ATEX certificate number Bas06ATEX0019X indicates that special conditions are required for safe operation in Europe These conditions are 1 The equipment must be protected against...

Страница 17: ...plied to the unit The transmitter requires up to 60 seconds to initialize and begin normal operation The unit will meet its specified accuracy within 3 minutes Figure 7 below shows a close up view of...

Страница 18: ...st temperature Node ID enter value Baud rate Parity Stop bits Data Bits Save Table RS485 Service Zero mV Zero PSIg Span mV Span PSIg Temp C 6 CS 6 Temp C 1 CS 1 Row 16 Units 16 Row 1 Units 1 Figure 8...

Страница 19: ...level passcode must be entered 2 2 3 Selecting measurement parameters To select the measurement parameters complete the following steps To select the number of measurements to be displayed on each scr...

Страница 20: ...he above screen highlight that choice and press enter Use the arrow keys to change the contrast value and press enter Then press escape twice to return to the main menu 2 2 5 Setting up the analog out...

Страница 21: ...twork when used with PanaView interface software To enter your nNode ID proceed as follows Enter your passcode to access the setup program Then use the arrow keys to select the Node ID menu Press ente...

Страница 22: ...o enter the desired pressure value and press enter The screen will be updated to show the new pressure value To set the temperature display enter your password and use the arrow keys to select const t...

Страница 23: ...to select lock keypad on the initial screen and press enter twice The keypad is now locked To unlock the keypad just re enter the user program as described on page 11 No further programming steps are...

Страница 24: ...creen instructions After completing the above step respond to the ARE YOU SURE question Be sure to respond yes if you wish to save the new information 3 2 Moisture probe error conditions IMPORTANT All...

Страница 25: ...al depends on the specific application To accomplish this either return the RTE to Panametrics for recalibration or install a new RTE The HygroPro electronics will automatically read and store the cal...

Страница 26: ...n set at 50 C 2 C 122 F 3 6 F for 24 hours 3 3 6 Installing the probe in the transmitter To install a new or cleaned probe in the transmitter complete the following steps 1 Reconnect the probe cable t...

Страница 27: ...iven a 24 VDC power supply Max load resistance 24 x 33 33 300 500 Cable 6 ft 2 m standard consult Panametrics for custom lengths Input parameters for loop powered intrinsic safety Ui 28V Pi 0 653W Li...

Страница 28: ...lable ranges 30 to 300 psig 3 to 21 bar 50 to 500 psig 4 to 35 bar 100 to 1000 psig 7 to 69 bar 300 to 3000 psig 21 to 207 bar 500 to 5000 psig 35 to 345 bar Note Psig ranges are based on a constant p...

Страница 29: ...s arising from course of dealing or usage or trade Return policy If a Panametrics instrument malfunctions within the warranty period the following procedure must be completed 1 Notify Panametrics givi...

Страница 30: ...uction targets across critical industries including Oil Gas Energy Healthcare Water and Wastewater Chemical Processing Food Beverage and many others Join the conversation and follow us on LinkedIn lin...

Отзывы: