Chapter 5. Specifications
5.1 Operation and performance
Fluid types
Acoustically conductive gases
Pipe sizes
2 to 16 in. (50 to 400 mm)
Meter body materials
Carbon Steel, SA216 Gr. WCB
Low Temp Carbon Steel, SA352 Gr. LCB
Stainless Steel, SA351 Gr. CF8M
Duplex Stainless Steel, SA995 Gr. CD3MWCuN
Flange ratings
ASME:
150 lb, 300 lb or 600 lb
DIN:
PN10, PN16, PN25, PN40 or PN63
Meter body certifications
PED Cat III, Module B+C2
CRN (All Canadian Provinces)
NACE MR01-75/MR-01-03
1
Flow measurement range (bi-directional)
0.5 to 250 ft/s (0.15 to 76 m/s)
Turndown Ratio = 500:1
Flow measurement accuracy
• 1 Path:
• ±0.075 ft/s (±0.02 m/s) at flow range of 0.5 to 5 ft/s
(0.15 to 1.5 m/s)
±1.5% of reading at flow range of 5 to Vmax/s (1.5 to
Vmax)
• 2 Path:
±0.05 ft/s (0.015 m/s) at flow range of 0.5 to 5 ft/s
(0.15 to 1.5m/s)
±1.0% of reading at flow range of 5 to Vmax (1.5 to 76
Vmax)
Note:
Accuracy/repeatability specifications assume a
final installation with fully developed flow profile (typically
20 diameters upstream and 10 diameters downstream of
straight pipe run), Reynolds Number> 5000 and single phase
fluids. Applications with piping arrangements that induce
swirl (e.g., two out-of-plane elbows) may require additional
straight run and/or flow conditioning. For shorter straight
pipe runs, consult the factory for a computational flow
dynamic evaluation.
Repeatability
±0.2% to 0.5% of reading
Measurement parameters
Mass flow, standard and actual flow, totalized flow, and flow
velocity
Electronics enclosure
NEMA Type 4X explosion-proof and weatherproof (IP66)
Standard:
Epoxy-coated aluminum
Optional:
Stainless steel
Hazardous area certifications
US/CAN:
Class 1, Div. 1 Group B,C,D
ATEX:
II 2 G Ex db IIB+H2 T6... 150° C Gb
IECEx:
Ex db IIB+H2 T6... 150° C Gb
Input power
Standard:
90 to 240 VAC
Optional:
12 to 28 VDC, ±5%
Power consumption
20 W maximum
Process temperature range
–40 to 302°F (–40 to 150°C)
Note:
-40 to 257° F (-40 to 125° C) range when used with
pressure and temperature sensor option.
Ambient temperature range
-40 to 140°F (–40 to 60°C) US/CAN
-4 to 140°F (–20 to 60°C) ATEX/IECEX
Below -20 Display readability will be affected
Storage temperature
-40 to +185°F (-40 to +85°C)
Digital communication
Standard:
RS485
Optional:
HART®
Optional:
Modbus®
Optional:
Foundation Fieldbus®
28
Содержание PanaFlow Z1G
Страница 1: ...PanaFlow Z1G Z2G User s manual 910 321 Rev A...
Страница 2: ...ii...
Страница 4: ...no content intended for this page iv...
Страница 9: ...no content intended for this page 1...
Страница 21: ...Figure 12 Remote mount electronics transducer and preamplifier wiring ref dwg 702 731 732 13...
Страница 28: ...no content intended for this page 20...
Страница 30: ...no content intended for this page 22...
Страница 38: ...no content intended for this page 30...
Страница 40: ...no content intended for this page 32...
Страница 43: ...Table 13 Service record cont Date Description of service Performed by 35...