Copyright 2003 Baker Hughes Company.
8 | DPI800/802 Instruction Manual–English
Selection Sequence:
Select Task (Table
Set Up (Table
Calibration
When you select a channel, the display shows the
applicable instructions to complete the calibration.
When the calibration is complete, select Calibration Due
and set the new calibration date for the instrument.
4.2 Procedures (Pressure P1/P2)
1. Connect the instrument to the pressure standard
2).
2. Let the equipment get to a stable temperature
(minimum: 30
minutes since the last power on).
3. Use the calibration menu (Table
6) to do a two
-
point
calibration (Zero and +FS) or a three-point calibration (-
FS, Zero and +FS). Refer to Table
7. The display
shows the applicable instructions to complete the
calibration.
4. To make sure that the calibration is correct, select the
applicable pressure task (Table
2) and apply these
pressure values:
•
Ranges g/d or sg: 0, 20, 40, 60, 80, 100 (%FS)
Then: Go back to 0 in the same steps.
Then (three-point calibration only):
-
20,
-
40,
-
60,
-
80,
-
100 (%FS).
Then: Go back to 0 in the same steps.
•
Ranges a: 0, 20, 40, 60, 80, 100 (%FS)
Then: Go back to 0 in the same steps.
5. Make sure that the error is in the specified limits:
•
Standard accuracy:
The specified accuracy (refer to “Specification
9) includes an allowance for
temperature changes, reading stability for one
year, and the uncertainty of the standard used for
calibration.
In step 4, make sure that the error between the
applied pressure and the reading on the unit is not
more than 0.015% FS.
•
Premier precision:
The specified precision (refer to “Specification
9) includes an allowance for
temperature changes and the uncertainty of the
standard used for calibration.
In step 4, make sure that the error between the
applied pressure and the reading on the unit is not
more than the specified value for Premier
precision.
4.3 Procedures (mA input)
1. DPI802/802P only. Connect the instrument to the
calibration equipment (Figure
7).
2. Let the equipment get to a stable temperature
(minimum:
5
minutes since the last power on).
3. Use the calibration menu (Table
6) to do a three-point
calibration (-FS, Zero and +FS). The display shows the
applicable instructions to complete the calibration.
4. To make sure that the calibration is correct, select the
applicable mA task (Table
2) and apply these values:
•
mA:
-
55,
-
40,
-
24,
-
18,
-
12,
-
6, 0 (open circuit)
Then: 0, 6, 12, 18, 24, 40, 55.
1
Display: Enter
Calibration PIN
2
3
Menu: Select
Channel
4
5
+
Factory PIN =
4321
6)
Table
6: Calibration Options
Options
(If applicable) Description
P…
To calibrate the specified IDOS pressure input
(P1
or
P2
a
).
a. Optional item
IDOS …
UMM only. To calibrate the specified IDOS
UMM. Refer to the user manual for the IDOS
UMM.
mA
DPI802/802P only. To calibrate the mA input.
Calibration Due: To set the date of the next
calibration for the instrument.
After the specified calibration date, there is a
warning message. There is a selection box to
stop the warning.
To change the calibration PIN (Personal
Identification Number).
Table
7: Calibration Pressures
Ranges: g/d
Nominal Applied Pressure
psi (mbar)
-FS
a
a. For a three-point calibration, do not apply more than
-
90% of the
specified FS for the unit.
Zero
+FS
≤ 10.0 psi (700 mbar)
-FS
0
+FS
> 10.0 psi (700 mbar)
-
13.1 (
-
900)
0
+FS
MENU
OK
4 3 2 1
MENU
OK
P1... g
MENU
OK
Ranges: a
Nominal Applied Pressure
psi (mbar)
Zero
+FS
5.00 psi (350 mbar)
< 0.02 (1.0)
+FS
30.0 psi (2 bar)
< 0.07 (5.0)
+FS
100.0 psi (7 bar)
< 0.29 (20.0)
+FS
300.0 psi (20 bar)
< 0.73 (50.0)
+FS
Ranges: sg
Nominal Applied Pressure
psi (mbar)
Zero
+FS
≥ 5000 psi (350 bar)
Use atmospheric
pressure as zero.
+FS
Содержание Druck Sensycal DPI800
Страница 2: ......
Страница 36: ...Copyright 2003 Baker Hughes Company 32 DPI800 802 Bedienungsanleitung Deutsch...
Страница 48: ...Copyright 2003 Baker Hughes Company 44 Manual de instrucciones DPI800 802 Espa ol...
Страница 60: ...Copyright 2003 Baker Hughes Company 56 DPI800 802 Notice d emploi Fran ais...
Страница 72: ...Copyright 2003 Baker Hughes Company 68 Manuale d uso DPI800 802 Italiano...
Страница 84: ...Copyright 2003 Baker Hughes Company 80 Manual de Instru es do DPI800 802 Portugu s...
Страница 119: ......