English–DPI800/802 Instruction Manual | 7
2.7 mA Measurements
DPI802/802P only. To measure a current:
1. Connect the instrument (Figure
7) and, if necessary,
adjust the Set Up (Table
1).
2. Select the applicable mA task from Select Task
3) and, if necessary, adjust the
Settings (Table
4).
Figure
7: Example Configuration - mA Measurement
2.8 UPM Pressure Measurements
Read all the instructions supplied with the UPM and then
use the specified procedures to connect it (Figure
8).
Figure
8: Example Configuration - Pressure
Measurement with a UPM
When the connections are complete, make the necessary
IDOS selections (Table
3). Then use the same
procedures as specified for the P1/P2 connections.
If you re-attach a UPM, the instrument uses the same
measurement units that you used before. The instrument
keeps a record for the last 10 modules.
2.9 Error Indications
If the display shows <<<< or >>>>:
•
Make sure that the range is correct.
•
Make sure that all the related equipment and
connections are serviceable.
3. Maintenance
This section gives procedures to maintain the unit in a good
condition. Return the instrument to the manufacturer or an
approved service agent for all repairs.
3.1 Clean the Unit
Clean the case with a moist, lint-free cloth and a weak
detergent. Do not use solvents or abrasive materials.
3.2 Replace the Batteries
To replace the batteries, refer to Figure
B1. Then re-attach
the cover.
Make sure that the time and date are correct. The
calibration facility uses the date to give service and
calibration messages.
All the other configuration options stay in memory.
3.3 Return Goods/Material Procedure
If the unit requires calibration or is unserviceable, return it
to the nearest Druck Service Centre listed at:
https://druck.com/service
.
Contact the Service Department to obtain a Return
Goods/Material Authorization (RGA or RMA). Provide the
following information for a RGA or RMA:
•
Product (e.g. DPI802)
•
Serial number.
•
Details of defect/work to be undertaken.
•
Calibration traceability requirements.
•
Operating conditions.
4. Calibration
Note:
Druck can provide a calibration service that is
traceable to international standards.
We recommend that you return the instrument to the
manufacturer or an approved service agent for calibration.
If you use an alternative calibration facility, make sure that
it uses these standards.
4.1 Before You Start
To do an accurate calibration, you must have:
•
The calibration equipment specified in Table
5.
•
A stable temperature environment: 70 ± 2°F (21 ± 1°C)
Before you start the calibration, make sure that the time
and date on the instrument are correct (Table
1).
30V MAX
30V MAX
mA
24V
30V MAX
30V MAX
mA
24V
MENU
OK
HOLD
ESC
mA
MENU
OK
mA
mA(24V)
mA(24V)
MENU
OK
-
+
+VE
-
+
-VE
+VE
-VE
0.083
55mA
0.083
55mA
Maximum: 30V
Settings
Settings
a) with external
loop power
b) with internal
loop power
IDOS ... g
MENU
OK
HOLD
ESC
IDOS
IDOS 30 ... g
30 ... g
MENU
OK
IDOS
30 ... g
MENU
OK
15.03
15.88
15.74
0.14
(@16:11:18)
- (@16:11:35)
=
2.76 ... /min
5.03
Settings
a) Pressure
b) Leak test
Start
Settings
Table
5: Calibration Equipment
Function
Calibration Equipment
Pressure
An applicable pressure standard (primary or
secondary) with a total uncertainty of 0.01%
reading or better.
mA
mA calibrator.
Accuracy: Refer to Table
Содержание Druck Sensycal DPI800
Страница 2: ......
Страница 36: ...Copyright 2003 Baker Hughes Company 32 DPI800 802 Bedienungsanleitung Deutsch...
Страница 48: ...Copyright 2003 Baker Hughes Company 44 Manual de instrucciones DPI800 802 Espa ol...
Страница 60: ...Copyright 2003 Baker Hughes Company 56 DPI800 802 Notice d emploi Fran ais...
Страница 72: ...Copyright 2003 Baker Hughes Company 68 Manuale d uso DPI800 802 Italiano...
Страница 84: ...Copyright 2003 Baker Hughes Company 80 Manual de Instru es do DPI800 802 Portugu s...
Страница 119: ......