4 | PACE Calibration Manual–English
4. Preliminary Operations
Review and become familiar with the whole procedure before performing a calibration.
Before performing a calibration:
1. Turn the PACE on and allow it to thermally stabilise (for at least 2 hours) in a thermally stable
environment.
2. Carry out a leak test as detailed in PACE User Manual K0443.
5. Notes on Calibration
The pressure calibration standard output port and the PACE reference level must be at the same
level. See illustrations below for PACE reference level. If the pressure calibration standard is not
at the PACE reference level, use height-corrected applied pressure.
Figure
1: PACE1000 Reference Level
Figure
2: PACE Control Module Reference Level
Set the PACE units of pressure to one of the required units for calibration.
1
Reference Level
1
Reference Level
1
1