Chapter
2.
Getting Started
2.3 Change Pressure Units
To change the pressure units proceed as follows. If the four units displayed are not the units
required, press
TASK
and select any task, other than
BASIC
, press
SETUP
and proceed as
detailed in Section
7.1. To return to
BASIC
mode, press
TASK
and select
BASIC
.
In
BASIC
mode, the unit is configured to carry out basic Pressure to Voltage (
P
to
V
) or Pressure
to Current (
P
to
I
) tests, a typical test procedure follows:
2.4 Voltage and Current Measurements
Connect the electrical input sockets as follows for voltage and current measurements. Use the
test leads provided and
DO NOT
push bare wires into the sockets.
Note:
See Table
2 for maximum instrument ratings.
SELECT UNITS
TASK : BASIC
VOLTAGE
PRESSURE INT
CURRENT
PRESSURE
UNITS
SELECT UNITS
OF PRESSURE
55mA
Vin
mA in
mA Sink
max 30V
max
+
-
V
55mA
Vin
mA in
mA Sink
max 30V
max
+
-
Содержание Druck DPI 610 IS
Страница 1: ...Druck com DPI 610 615 IS Intrinsically Safe Portable Pressure Calibrator Instruction Manual...
Страница 2: ......
Страница 6: ...Copyright 2007 Baker Hughes Company iv DPI 610 615 IS Instruction Manual English...
Страница 10: ...Copyright 2007 Baker Hughes Company viii DPI 610 615 IS Instruction Manual English...
Страница 20: ...Copyright 2007 Baker Hughes Company 10 DPI 610 615 IS Instruction Manual English Chapter 2 Getting Started...
Страница 38: ...Copyright 2007 Baker Hughes Company 28 DPI 610 615 IS Instruction Manual English Chapter 5 Advanced Task...
Страница 64: ...Copyright 2007 Baker Hughes Company 54 DPI 610 615 IS Instruction Manual English Chapter 8 Calibration...
Страница 75: ......