Table 1-6. Type PTSPA Absolute Pressure Transmitter Specifications
Property
Characteristic/Value
Measurement range,
turndown ratio, zero
suppression, and zero
elevation
Lower range value (zero) and upper range value (100%) can be calibrated at any
value of pressure, provided that:
1. Their algebraic difference (the calibrated span) corresponds to an
authorized TDR.
2. Both are within the following acceptable limits.
Stability
±0.1% of URL per 12-month period (at reference conditions)
Reference accuracy -
per IEC 770 and SAMA
PMC 31.1
The reference accuracy includes the effects of linearity, hysteresis, repeatability,
and deadband. These values are for transmitters with zero based spans, silicone
oil fill, and Hastelloy C-276 isolating diaphragms at a reference temperature of
25°C (77°F). The value listed may vary depending on the URL and calibrated span
of the particular transmitter.
PTSPAE
±0.20% of span or ±
(% of span), whichever is greater
PTSPAG and
PTSPAJ
±0.10% of span or ±
(% of span), whichever is greater
Pressure limits
The minimum sustained absolute pressure is 3.4 kPa (0.5 psi). The maximum
overpressure is listed in the following table.
Overpressure effect
This value is for transmitters with zero based spans, silicone oil fill, and
Hastelloy C-276 isolating diaphragms at a reference temperature of 25°C (77°F).
±0.05% of URL for line pressures 1.5 times URL.
Ambient temperature
effect on zero and span
These values are for transmitters with zero based spans, silicone oil fill, and Has-
telloy C-276 isolating diaphragms. The effects listed are for a variation of ±25°C
(±45°F) from a calibration done at a temperature between -30° and +80°C (-22°
and +176°F), as long as the variation does not take the transmitter out of the tem-
perature specifications listed in Table
.
PTSPAE
±(0.10% URL + 0.10% span)
PTSPAG and PTSPAJ ±(0.08% URL + 0.10% span)
Weight
1.3 kg (2.9 lbs) without options or accessories.
SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE.
Type
International System Units
Range Limits (kPa)
Span and Turndown Ratio (TDR)
Upper
Lower
Nominal (kPa)
Minimum (kPa)
PTSPAE
200
0
200
1:1 TDR
5
40:1 TDR
PTSPAG
3,000
0
3,000
1:1 TDR
75
40:1 TDR
PTSPAJ
40,000
0
40,000 1:1 TDR
1,000
40:1 TDR
Type
American Units
Range Limits (psi)
Span and Turndown Ratio (TDR)
Upper
Lower
Nominal (psi)
Minimum (psi)
PTSPAE
30
0
30
1:1 TDR
0.8
40:1 TDR
PTSPAG
450
0
450
1:1 TDR
11
40:1 TDR
PTSPAJ
5,800
0
5,800
1:1 TDR
145
40:1 TDR
0.016
URL
×
Span
--------------------------------
0.013
URL
×
Span
--------------------------------
Type
URL
Overpressure Limit
(kPa)
(psi)
(kPa)
(psi)
PTSPAE
200
30
600
87
PTSPAG
3,000
450
6,000
870
PTSPAJ
40,000
5,800
41,368
6,000