Microbial  contamination  of  reagents  while  taking  aliquots  should  be  avoided.  The  use  of  sterile 
disposable  pipettes  and  pipette  tips  is  recommended.  Do  not  use  reagents  looking  cloudy  or 
showing signs of microbial contamination. 

 

A Material Safety Data Sheet (MSDS), respectively a declaration on Material Safety Data Sheets is 
available for download at 

www.bag-diagnostics.com

 .

 

 
 

8. 

KIT SPECIFICATIONS 

The combination of primers and probes allows the determination of human HLA class I and II alleles 
according to lot-specific data (low to medium resolution, detection of all alleles with the exception of 
single  rare  alleles).  The  accuracy  and  reproducibility  of  the  reactivity  of  the  test  kit  is  checked  for 
each lot  with control specimens with known HLA-alleles. The kit determines the HLA-Loci A, B, C, 
DRB1, DRB3, DRB4, DRB5, DQA1, DQB1, DPA1 and DPB1.

 

 
 

8.1  SPECIFIC PERFORMANCE CHARACTERISTICS 

 

8.1.1  Diagnostic sensitivity and specificity 

 

A total of 116 DNA samples pre-typed for the relevant HLA loci or international reference cell lines 
were  tested  with  the  HISTO  TYPE  Rainbow  QS6  kit  to  show  the  correct  reactivity.  There  was  a 
100% concordance of results with the pre-typings for all loci.

 

 

Locus 

Number of sample pre-typed for the locus 

concordant 

% concordance 

116 

116 

100% 

116 

116 

100% 

116 

116 

100% 

DRB1 

115 

115 

100% 

DRB3 

113 

113 

100% 

DRB4 

113 

113 

100% 

DRB5 

113 

113 

100% 

DQA1 

114 

114 

100% 

DQB1 

114 

114 

100% 

DPA1 

113 

113 

100% 

DPB1 

114 

114 

100% 

 

 

8.1.2  Analytical sensitivity and measuring range 

 

Validation tests have shown that variation of the amount of DNA from 8 ng to 30 ng per reaction has 
no significant effect on the specific detection of HLA alleles. 
 
 

 

Содержание HISTO TYPE Rainbow QS6

Страница 1: ...ample information into PlexTyper Software 5 6 4 Amplification 5 6 5 Export of results from QuantStudio 6 Flex System 9 6 6 Evaluation and interpretation of the results 10 6 7 Import of the results fil...

Страница 2: ...s amplified in a real time PCR with sequence specific primers SSP The primers were specially developed for the selective amplification of segments of specific HLA alleles or allele groups The amplicon...

Страница 3: ...2 Real time PCR Cycler validated cycler see 4 3 Plate holder QS6 REF 726321 Variable pipettes 0 5 1000 l and pipette tips Application spatula for qPCR Seal Molecular grade DNAse free water Suitable pl...

Страница 4: ...amplification detection use two separate rooms if possible Use devices and other materials only at the respective workplaces and do not exchange them 6 2 DNA Isolation The specimen material for the i...

Страница 5: ...entity This Run ID should be used when setting up the file identity in the PCR machine If the PCR excel export file is not pre labelled with the RUN ID the file can be manually prefixed with the corre...

Страница 6: ...te the test DNA with other concentrations For DNA of different stock concentrations use the dilution table below to adjust the DNA to working concentration Add 8 l Molecular grade water and 2 l Plex M...

Страница 7: ...ely sealed particularly at the edge of the plate Make sure that the liquid has contacted the dried mix and there are no bubbles or air gaps in the reaction wells If bubbles appear gently tap the tubes...

Страница 8: ...Flex System Block type Fast 96 Well 0 1mL Experiment type Comparative Ct Ct Reagent type TaqMan Reagents Run properties Standard Define Targets Please note Custom dye instrument calibration must be p...

Страница 9: ...rt the raw data in excel format with the following steps 1 Select the eds file double click on a eds file If the QuantStudio RealTime PCR software is installed the software will automatically open the...

Страница 10: ...s required for the evaluation are available from the BAG Diagnostics download server www service bag diagnostics com Please make a note of the lot number of the kit The interpretation kit files are pr...

Страница 11: ...s possible contamination by the PlexTyper software and a warning message is generated Amplification signals above Cq 36 in the NTC are regarded as PCR artefacts and are disregarded If PCR contaminatio...

Страница 12: ...No results at an HLA locus 7 WARNINGS AND DISPOSAL INSTRUCTIONS HISTO TYPE Rainbow QS6 is designed for in vitro diagnostic use The kit should only be used by specially trained qualified personnel All...

Страница 13: ...t is checked for each lot with control specimens with known HLA alleles The kit determines the HLA Loci A B C DRB1 DRB3 DRB4 DRB5 DQA1 DQB1 DPA1 and DPB1 8 1 SPECIFIC PERFORMANCE CHARACTERISTICS 8 1 1...

Страница 14: ...n to avoid contamination of kit reagents and other laboratory materials with amplicons or DNA The performance of a negative control without DNA in well H12 is strongly recommended No fluorescence sign...

Страница 15: ...ll of the tube Careful pipetting Spin down PCR plate User error Ensure all wells receive the required volume of reagents Evaporation of the reagents due to incorrect closing of the PCR tubes Make sure...

Страница 16: ...ue number RTU Ready to use 14 LITERATURE 1 Mack S J et al 2013 Tissue Antigens 81 194 203 2 Beutler E et al 1990 BioTechniques 9 166 15 REVIEW HISTORY Version 01 2021 Issue 2021 04 Initial version Ver...

Отзывы: