Additional programming: Industrial analog and industrial pulse 

Page 10/15 

 

 

AC_ILR_BA_02_1805 

 

4.4.  Linearisation 

(for models ILR701, ILR701T, ILR750 and ILR750T) 

 

Indicated by 1 – 9 on the left hand side of the display, followed by a hyphen (-), this 

screen allows the setting of the linearisation (in total 9 points). 

 

 

 

 

 

 

 

 

 

Press  the 

TOTAL

  button  to  select  a  digit  (selected  digits  flash).  Press 

RESET

  to 

change the selected digit. The flow rate will be set in the top row of the meter and is 

displayed in the unit you selected at step 9.1 (unit of measure). In the sample shown 

above this would be the flow rate 0.4 liter per minute. On the bottom line of the meter 

you can set in the correction of the error in %. In the sample below, the error at a flow 

rate of 0,4 liter per minute would be -7,82%; to correct this, +7,82% needs to be set in 

(the plus symbol [+] will not be shown). 

Once  the  adjustment  of  the  linearisation  is  completed,  press  and  hold  the 

TOTAL

 

button for one second to advance to the next linearisation point. 

 

 

 

 

 

 

 

 

Number 9 at the left hand side of the display shows the 9

th

  linearisation point. The 

sample shows a flow rate of 250.0 liter per minute and a deviation of the flow meter of 

+0,15%. To correct this error, -0,15% needs to be set as correction. 

 

Note:  

- Minimum 3 linearisation points needs to be programmed. 

- The flow rates do not have to be programmed from low to high; the software will 

sort  the  flow  rates  automatically,  no  matter  at  which  point  (1-9)  they  are 

programmed. 

 

To exit the programming mode: 

On any screen, press and hold both the 

TOTAL

 and 

RESET

 buttons. The screen 

will revert back to the programmed scale factor, and then flash. Following the three 

flashes, the register display will be blank. 

 

Note:

 Pressing the 

TOTAL

 or 

RESET

 buttons will turn the display back on.  

Figure 13: Linearisation point 1 (of 9) 

Figure 14: Linearisation point 9 (of 9) 

 

Содержание ILR710

Страница 1: ...Badger Meter Europa GmbH Registers type ILR7XX and ILR7XXT INSTALLATION AND OPERATION MANUAL May 2018 AC_ILR_BA_02_1805...

Страница 2: ...s with display 6 3 5 Changing the display mode for all ILR models with display 7 3 6 Exiting programming mode for all ILR models with display 7 4 Additional programming Industrial analog and industria...

Страница 3: ...ectric shock Failure to properly earth a unit may cause damage to that unit or data stored within it Protection class The device has protection class IP 65 and needs to be protected against dripping w...

Страница 4: ...mmendations Page 2 15 AC_ILR_BA_02_1805 RoHs Our products are RoHs compliant Battery disposal The batteries contained in our products need to be disposed of as per your local legislation acc to EU dir...

Страница 5: ...press either the TOTAL or RESET button once 2 2 Status The status indicators are RESET and TOTAL 2 3 Totalizers The top row of indicators is the batch totalizer This totalizer displays the cumulative...

Страница 6: ...are calculated in volume unit per minute 2 5 Battery The LBat indicator will indicate when the battery is approaching end of life When the indicator is illuminated the 2 3AA 3 0 VDC lithium battery i...

Страница 7: ...ale factor programming 3 2 Scale factor for all ILR models with display The register collects input pulses from the oval gear meter and then determines the appropriate display output using the scale f...

Страница 8: ...9999 99 on the top row integer values are displayed on the bottom row Example 10 45 would display 45 on the top row and 10 would be displayed on the bottom row 2 When finished adjusting the pulse rate...

Страница 9: ...le a letter is flashing on the display press the RESET button to select either totalizer or flow rate 2 Upon completion of this setting the programming of the industrial standard register and the indu...

Страница 10: ...s The Output Pulse Length should be set to less than the value of t Per the equation Maximum meter flow rate in GPM or l m t x 1000 60X output pulse rate where t the required pulse rate in millisecond...

Страница 11: ...o advance to the next screen To advance to the next programming screen hold the TOTAL button for one second 4 3 Analog maximum flow rate for models ILR750 and ILR750T Indicated by a H on the left hand...

Страница 12: ...he plus symbol will not be shown Once the adjustment of the linearisation is completed press and hold the TOTAL button for one second to advance to the next linearisation point Number 9 at the left ha...

Страница 13: ...Yellow External ground Brown Pulse output White DC Input 8 to 24 VDC 20 to 40mA Outputs Pulse output with internal pull up resistor optional open collector output with output jumper removal pulse out...

Страница 14: ...nput 8 to 24 VDC 20 to 40mA Outputs Dual pulse output with internal pull up resistor optional open collector output with output jumper removed dual pulse output forms a quadrature signal for direction...

Страница 15: ...ngs Max power 10W not to exceed max voltage 200 VDC peak AC max current 0 5A DC peak AC Outputs Raw reed switch output with no signal conditioning Pulse per unit of measure IOG series Note Actual puls...

Страница 16: ...DC 20 to 40mA Outputs Analog 4 to 20mA output in loop powered configuration external load of 50 ohms to 250 ohms flow rate is linear scaled between 4mA minimum and 20mA maximum set points flow rates b...

Страница 17: ...ir Harmlessness declaration Page 15 15 AC_ILR_BA_02_1805 6 Return of goods for repair Harmlessness declaration Please refer to our claims return form harmlessness declaration under www badgermeter de...

Страница 18: ...line Phone 49 7025 9208 0 or 46 Fax 49 7025 9208 15 Badger Meter Europa GmbH Subsidiary of Badger Meter Inc USA N rtinger Strasse 76 72639 Neuffen Germany E mail badger badgermeter de www badgermeter...

Отзывы: