Pulse Output Programming
The
Pulse Output
can be programmed for any engineering unit of measure . A
Pulse Output Scale Factor
must be calculated and
programmed (
Program Mode 6
) using the same formula and procedure as described under the
Totalizer Scale Factor
(
Program
Mode 1
) .
The transmitter pulses per unit in Table 3 apply to the following transmitters:
• FT-1
• FT-1E
• FT-2
• PFT-2
• PFT-2E
• PFT-3
• FT-420
• PFT-420
Size
(in.)
Meter Model
US Gallons
Liters
Ft
3
1/2
OP
222 .960
58 .899
1687 .857
1/2
OP (FT1 only)
111 .500
29 .455
834 .078
1
OP
76 .640
20 .246
573 .307
2
OP
20 .600
5 .442
154 .099
2
Industrial Turbo
17 .360
4 .586
129 .862
3
Industrial Turbo
12 .400
3 .276
92 .758
4
Industrial Turbo
2 .560
0 .676
19 .150
6
Industrial Turbo
1 .080
0 .285
8 .079
5/8
LP RCDL
229 .554
60 .641
1717 .186
5/8
25 RCDL
198 .340
52 .396
1484 .689
3/4
35 RCDL
126 .671
33 .463
947 .566
1
40 RCDL
89 .781
23 .717
671 .610
1
55 RCDL
58 .064
15 .339
434 .351
1
70 RCDL
46 .752
12 .350
349 .726
1-1/2
120 RCDL
23 .867
6 .305
178 .539
2
170 RCDL
14 .565
3 .848
108 .955
1-1/2
160 Turbo Series
1 .537
0 .406
11 .495
2
200 Turbo Series
1 .537
0 .406
11 .495
3
450 Turbo Series
1 .598
0 .422
11 .955
4
1000 Turbo Series
1 .665
0 .440
12 .455
6
2000 Turbo Series
0 .150
0 .040
1 .122
8
3500 Turbo Series
0 .151
0 .040
1 .133
10
5500 Turbo Series
0 .198
0 .052
1 .481
12
6200 Turbo Series
0 .129
0 .034
0 .963
16
6600 Turbo Series
0 .016
0 .004
0 .116
20
10000 Turbo Series
0 .009
0 .002
0 .067
1/2
Industrial OG
378 .5 *
100 *
2831 *
3/4
Industrial OG
249 .8 *
66 *
1868 *
1
Industrial OG
249 .8 *
66 *
1868 *
1HF
Industrial OG
162 .8 *
43 *
1217 *
1-1/2
Industrial OG
64 .4 *
17 *
481 *
2
Industrial OG
34 .1 *
9 *
255 *
3
Industrial OG
11 .4 *
3 *
85 *
* If using a PFT-3E transmitter, multiply number by 2 .
Table 3: Transmitter pulses per unit
User Manual
Page 11
August 2018
REG-UM-00836-EN-12