POOGOAMMING

Each turbine flow meter is shipped with either a K-factor value or frequency data  If frequency data is provided, the data must 

be converted to a K-factor before programming the monitor  K-factor information, when supplied, can usually be found on the 

neck of the flow meter or stamped on the flow meter body  The K-factor represents the number of pulses per unit of volume  

The K-factor will be needed to program the monitor readout 

Programming Mode

x

x

See Note

Note: Only available when either

          KGS or LBS are Selected

Figure 4:  Advanced programming menu

Enter Programming Mode

Press 

MENU

 to enter the programming mode  The mode indicator changes from 

RUN

 to 

PROGRAM

 

OTEE:

N

If any input value exceeds the meter’s capabilities for that particular parameter, the 

LIMIT

 indicator flashes, indicating 

an invalid entry  Press 

ENTEO

 to return to the parameter’s entry screen to reenter the value 

Select Meter Size

At the 

METER

 prompt, press 

ENTEO

  The current meter size flashes  If the current setting is correct, press 

ENTEO

 to advance to 

the next parameter  To change the current setting, press 

UP

 or 

OIGHT

 to scroll to the bore size of your meter  Press 

ENTEO

 to 

save and advance to the 

DISPLAY

 parameter 

OTEE:

N

The meter connection size and the bore size are different  For example, many of the one inch NPT turbines have bore 

sizes that range from 3/8…1 in  (9 53…25 4 mm)  Use the correct bore size so the meter will report correct flows and 

totals 

Select Display Function

The monitor can display 

RATE

 or 

TOTAL

 or alternate between 

BOTH

 rate and total  At the 

DISPLAY

 prompt, press 

ENTEO

  The 

monitor shows the current display mode  To change to an alternate display mode, press 

UP

 or 

OIGHT

 to scroll to the correct 

display mode  Press 

ENTEO

 to save and advance to the 

RATE UNT

 parameter 

Select Units of Measure

At the 

RATE UNT

 prompt, press 

ENTEO

  The monitor shows the current rate unit  If the current selection is correct, press 

ENTEO

 

to advance to the next parameter  To change to an alternate unit, press 

UP

 or 

OIGHT

 to scroll to the correct rate unit  Press 

ENTEO

 to save and advance to the 

RATE INT

 parameter 

Programming

Page 8 

March 2017 

DSY-PM-00279-EN-03

Содержание Blancett B2800XP

Страница 1: ...Flow Monitor B2800XP Explosion Proof Flow Monitor Advanced DSY PM 00279 EN 03 March 2017 User Manual...

Страница 2: ...Mode 11 Additonal Scaling Parameters 12 Flow 4 mA Setting 12 Flow 20 mA Setting 12 4 20 mA Calibration 12 4 20 mA Test 12 Additional Input Options 13 Battery Replacement 13 Maintenance 14 Additional...

Страница 3: ...serious personal injury Indicates a hazardous situation which if not avoided could result in severe personal injury or death Indicates a hazardous situation which if not avoided is estimated to be ca...

Страница 4: ...lectable units viewing in run mode and prompts for variables in program mode The display can show rate total or alternate between both rate and total MENU ENTER Programming Buttons Figure 1 B2800XP fl...

Страница 5: ...amage to the cover threads and glass lens when glass lens cover is used 5 Pull wires into enclosure making certain they are long enough to make the required connections and to remove the instrument or...

Страница 6: ...b Size Kit Number Meter Sizes 1 2 in NPT Hub B280 742 3 8 in 1 2 in 3 4 in with 1 2 in NPT End Fittings 1 in NPT Hub B280 737 All Sizes with 1 in NPT End Fittings and Larger Installation Kit P N B280...

Страница 7: ...ion turn on the lock out feature to prevent unauthorized access or changing the meter s setup parameters MENU ENTER Figure 3 Front panel Buttons MENU Enters and exits PROGRAM mode UP Scrolls forward t...

Страница 8: ...If the current setting is correct press ENTER to advance to the next parameter To change the current setting press UP or RIGHT to scroll to the bore size of your meter Press ENTER to save and advance...

Страница 9: ...rams were chosen as the RATE UNT or TOTL UNT the monitor advances to the SP GRAV parameter Specific Gravity At the SP GRAV parameter press ENTER The first digit of the current specific gravity flashes...

Страница 10: ...plitude of the pulse is dependent on the voltage level of the supply connected to the pulse output and is limited to a maximum 30V DC Linearization Linearization enhances the accuracy of the display T...

Страница 11: ...password and return to the METER parameter The reset pasword is now required to reset the totals on the monitor NOTE N Entering a password in the PASSWORD screen and leaving the default password in t...

Страница 12: ...4 20 mA Calibration The 4 20 mA Calibration parameter is the fine adjustment of the 4 20 mA output The 4 mA setting is typically between 35 50 To set the 4 mA value connect an ammeter in series with...

Страница 13: ...Power Figure 7 Circuit board layouts BATTERY REPLACEMENT Battery powered monitors use a single 1 5V D size alkaline battery When replacement is necessary use a clean fresh battery for continued troub...

Страница 14: ...ates some form of alternating signal that can be raw that is no format amplification or wave shaping can be done prior to transmission to the readout The output pulse rate is related proportionally to...

Страница 15: ...t 2 25 Pulse Output Signals for Non linear Processes Few flow meters actually behave in a linear way There is always some uncertainty about the exact flow at a given reported input value For many comm...

Страница 16: ...nitor pig tail This should stop the noise If not then try reorientating the meter to a new location Flow rate indicator bounces This usually indicates a weak signal Replace pickup and or check all con...

Страница 17: ...ity 30 mV p p Over Voltage Protected 30V DC Frequency Measurement Accuracy 0 1 Temperature Drift 50 ppm C maximum Optional Analog Output Loop Powered Version Type 4 20 mA current loop Resolution 1 400...

Страница 18: ...RT NUMBER INFORMATION NOTE Advanced B2800 monitors only The default is gallons per minute Program Level A Advanced B 28 A X X X X Power Option B Battery L Loop Mounting Style X Explosion proof Units o...

Страница 19: ...INTENTIONAL BLANK PAGE Page 19 March 2017 DSY PM 00279 EN 03...

Страница 20: ...Colonia Guadalupe Inn CP 01050 M xico DF M xico 52 55 5662 0882 Europe Eastern Europe Branch Office for Poland Latvia Lithuania Estonia Ukraine Belarus Badger Meter Europe ul Korfantego 6 44 193 Knur...

Отзывы: