POOGOAMMING
Each turbine flow meter is shipped with either a K-factor value or frequency data If frequency data is provided, the data must
be converted to a K-factor before programming the monitor K-factor information, when supplied, can usually be found on the
neck of the flow meter or stamped on the flow meter body The K-factor represents the number of pulses per unit of volume
The K-factor will be needed to program the monitor readout
Programming Mode
x
x
See Note
Note: Only available when either
KGS or LBS are Selected
Figure 4: Advanced programming menu
Enter Programming Mode
Press
MENU
to enter the programming mode The mode indicator changes from
RUN
to
PROGRAM
OTEE:
N
If any input value exceeds the meter’s capabilities for that particular parameter, the
LIMIT
indicator flashes, indicating
an invalid entry Press
ENTEO
to return to the parameter’s entry screen to reenter the value
Select Meter Size
At the
METER
prompt, press
ENTEO
The current meter size flashes If the current setting is correct, press
ENTEO
to advance to
the next parameter To change the current setting, press
UP
or
OIGHT
to scroll to the bore size of your meter Press
ENTEO
to
save and advance to the
DISPLAY
parameter
OTEE:
N
The meter connection size and the bore size are different For example, many of the one inch NPT turbines have bore
sizes that range from 3/8…1 in (9 53…25 4 mm) Use the correct bore size so the meter will report correct flows and
totals
Select Display Function
The monitor can display
RATE
or
TOTAL
or alternate between
BOTH
rate and total At the
DISPLAY
prompt, press
ENTEO
The
monitor shows the current display mode To change to an alternate display mode, press
UP
or
OIGHT
to scroll to the correct
display mode Press
ENTEO
to save and advance to the
RATE UNT
parameter
Select Units of Measure
At the
RATE UNT
prompt, press
ENTEO
The monitor shows the current rate unit If the current selection is correct, press
ENTEO
to advance to the next parameter To change to an alternate unit, press
UP
or
OIGHT
to scroll to the correct rate unit Press
ENTEO
to save and advance to the
RATE INT
parameter
Programming
Page 8