Appendix C
–
Factory Configuration Settings
Manual Part Number:
5-06-4900-53-0
Revision Level: C
ECO 044952
Revision Date:
May 08, 2018
Page
5-4
5.3
APPENDIX C
–
ANALYZER FACTORY CONFIGURATION
SETTINGS
Supply Voltage
90
– 250 VAC, 50-60HZ
12-36 VDC
Sensor Type
GP Sensor (Part No: 8-01-1000-02-3)
CAG-250 Nylon (Part No: C1-16-1000-01-1)
IT (CR) Sensor (Part No: 6-01-1001-91-0)
Safety Barrier Type
MTL (Passive)
STAHL (Passive)
PEPPERL & FUCHS (Active)
None
Alarm and Relay Setup Information
ICS And O
2
Alarm Relays Failsafe/Non-Failsafe:
FAIL-SAFE
O
2
Alarm Relay Ascending/Descending:
ASCENDING
O
2
Alarm Trigger Level (In %):
8%
High ICS (Start Inert) Level (In %):
7%
Low ICS (Stop Inert) Level (In %):
6%
Low Sensor Output Fault Level (In %):
1%
Inert on Fault Conditions Setup Information
Inert on Flow Fault Condition (Yes/No):
YES
Inert on Sensor Disconnect Fault Condition (Yes/No):
YES
Inert on Analog To Digital Converter Failure Fault Condition (Yes/No):
YES
Inert on Low Sensor Output Fault Condition (Yes/No):
YES
Analog Voltage Output Setting
0
– 1 Volts
X
0
– 5 Volts
0
– 10 Volts
Analog Current Output Setting
0
– 20 mA
4
– 20 mA
X