FULL

MEDIUM

LOW

REPLACE

IMPORTANT:

 Perform power-up and initialization in 

fresh air to ensure proper calibration.

1.

  Perform a visual inspection of the analyzer, sample probe 

and hose assemblies.

  Check for signs of splits and / or cracks in the sample tubing.

  Assemble, clean and dry the water trap.

2.

  Press Power Key to initiate 60-second warm-up.

  Address / acknowledge any faults or errors discovered 

during initialization.

  Ensure batteries have sufficient charge before operation.

WARNING

:

 This analyzer is not intended to be used as a 

safety device.

WARNING

:

 This analyzer has not been designed to be 

intrinsically safe for use in areas classified as hazardous 

locations.

CAUTION

:

 Perform routine maintenance (replacing sensors, 

calibrating, etc.) as outlined in the user manual.

IMPORTANT: 

Comply with all local and national laws, rules 

and regulations associated with this equipment. Operators 

should be aware of the regulations and standards in their 

industry / country for the operation of this analyzer.

8

T-Stack Port

9

USB Port

10

Sample Gas Port

11

Water Trap / Filter 

Assembly

12

Standard Probe

*

1

IrDA Port

2

Function Keys

3

Navigation Keys

4

Enter Key

5

Run / Hold Key

6

Escape Key

7

Power / Backlight Key

*

Optional; may not be included 

with all product configurations.

Combustion & Emissions

P/N: 0019-9374

 | December 2019 Revision 0

Quick Start 

Guide

High Range CO Analyzer

for Commercial & Light Industrial Use

3.  Component Overview 

1.  Introduction

The Monoxor

®

 XR is a portable exhaust gas analyzer designed to 

measure high CO concentrations discharged from forklifts and other 

combustion engines burning propane, gasoline, diesel, LPG and CNG.

Please read and strictly follow the contents of this document for 

information on the operation and maintenance of the analyzer.

For more detailed product information, scan here or 

visit 

www.mybacharach.com

 to access the Monoxor

®

 

XR User Manual 

(P/N 0019-9376)

.

2.  Safety Instructions

4.  Hardware Setup

3.

  Firmly  connect  sample  tubing  and  thermocouple  to  the 

appropriate ports at the analyzer’s base.