5.4.3. CO Sensor Replacement Procedure
Follow the procedure below for CO sensor replacement.
Step CO Sensor Replacement Procedure
1
Remove battery door and the
connector tubing from the CO sensor.
2
Remove CO cap by twisting counter
clockwise.
3
Gently pull CO sensor out of its socket.
4
Properly dispose of the old CO sensor.
5
Plug new CO sensor into its socket.
6
Install the CO cap by aligning it toward
the “open” position (12 o’clock) as
shown in the diagram below, then
twisting the cap clockwise
approximately 40
°
to the "closed"
position (2 o’clock).
7
Reattach tubing.
8
Calibrate the CO sensor using either
the standard calibration procedure or
the B-SMART
procedure
5.4.4. B-SMART
CO Sensor Replacement
Step
B-SMART
CO Sensor Replacement
1
Enter the CALIBRATION MENU. Note that this requires password validation
(see page 28).
2
Use the UP (
) and DOWN (
) arrow buttons to select B-Smart. Press
ENTER to display the B-Smart code screen.
3
Use the UP (
) and DOWN (
) arrow buttons to enter the 10-digit
alphanumeric code supplied with the pre-calibrated B-SMART
sensor. Use
the LEFT (
) and RIGHT (
) arrow buttons to move the cursor across the
screen. Press ENTER.