Configuration
Monoxor
Plus Manual
26
0019-9370 Rev 1
Setup Menu
Function
Provides a list from which the user may select the
desired method for zeroing the CO sensor.
•
Auto-Zero happens automatically at warm-up.
•
Manual zero is used to initiate the zeroing process
whenever desired.
Use the UP (
) and DOWN (
) arrow buttons to
highlight the desired zeroing method.
Press ENTER to save. Press ESC to quit without saving.
CO AUTO ZERO
CO MANUAL ZERO
By default, the Monoxor
Plus
automatically zeroes all
sensors on ambient air when the instrument is turned
on.
The Monoxor
Plus
can be set to perform and store a
manual zero for the CO sensor. The instrument uses
the stored value to indicate background CO values
after warm-up instead of performing an auto-zero on
the background gas.