background image

 

 

 

GDX-350 Transmitter 

Instructions for Replacing 

an Uncalibrated Sensor 

 

 

 

 

 

Instruction:  5600-9003 

Revision: 0 

September 13, 2012 

 

 

 

Replacement Instructions 

 

 

IMPORTANT:

  This sensor requires calibration before being placed into service. 

 

1.

 

Ensure  the area is safe for maintenance to be performed in accordance with area 

requirements and local regulations. 

2.

 

Remove power from the GDX-350 transmitter. 

3.

 

Loosen, unscrew, and remove the sensor head cap using the sensor replacement tool (P/N: 

5600-0002) as shown below. 

 

 

Figure 1.  Sensor Head Cap and Sensor Replacement Tool 

4.

 

Remove the old sensor assembly (sensor and printed circuit board), noting the position of 

the notch found on the PCB header. 

5.

 

Install the new replacement sensor module. 