Instruction 24-9435
10
Fyrite Tech
3.5.3 CO Channel
Setup
(Model 60 only)
The CO Channel Setup Screen is labeled “CO.”
1. Enter the Setup Mode per Section 3.5.1. If necessary, repeatedly
press the
ENTER
button until “CO” is displayed.
2. Press the
▲
or
▼
button until the desired parameter is dis played.
- no
Sensor disabled. Use this function to turn off the CO chan-
nel if the sensor is bad or missing to avoid the CO sensor
error message from appearing.
- 2Ero
Manual Zero. When the CO channel is set to man u al zero,
the analyzer does not zero the CO sensor to ambient condi-
tions during start up.
- A2
Auto Zero. With the CO channel set to auto zero, the CO
sensor is zeroed to the ambient CO level during start up.
Important:
When using this mode the analyzer must be
turned ON in fresh air, otherwise in cor rect CO read-
ings will occur.
3. Press
ENTER
to move to the next Setup Screen, or press
I/O
to exit
setup.
3.5.4 Exiting the Setup Mode
Press the
I/O
button at any time to exit the Setup Mode and turn OFF the
analyzer. Note that the last displayed parameter is automatically saved in
memory.