background image

EN   

Instruction for use

PRO-39

Automatic Blood Pressure Monitor

EN

1. INTRODUCTION

Thank you for purchasing the B.Well Wrist Blood Pressure Monitor 

PRO-39. Designed for convenient and easy operation, this device 

provides fast and reliable measurement of systolic and diastolic 

blood pressure as well as heart rate using the oscillometric 

measurement method.

The PRO-39 is a fully automatic, digital, wrist blood pressure 

measuring device.

Intended use:

 The Blood Pressure Monitor PRO-39 including 

accessories is a non-invasive blood pressure measurement system 

intended to measure the diastolic and systolic blood pressures 

and pulse rate of an adult individual by using the oscillometric 

measurement method.

Application area:

 for use by medical professionals or at home.

Important advantages of PRO-39:

  Up-to-date IntellectClassic technology uses oscillometric 

measurement during deflation for quick, precise and painless 

result.

  The Pulse Arrhythmia Detection technology.

  Last measurement memory.

  This device has been proven in clinical studies to provide 

excellent accuracy.

2. CLASSIFICATION OF BLOOD PRESSURE VALUES

Table for classifying blood pressure values (mmHg) according

to European Society of Hypertension (ESH)

Range

Systolic 

blood 

pressure

Diastolic 

blood 

pressure

Measures

Grade 3: severe 

hypertension

Higher or 

equal to 

180

Higher or 

equal to 110

Urgently seek 

medical advice!

Grade 2: moderate 

hypertension

160-179

100-109

Consult your doctor 

immediately

Grade 1: mild 

hypertension

140-159

90-99

Consult your doctor

High normal

130-139

85-89

Consult your doctor

Normal

Lower than 

130

Lower than 

85

Self-check

Optimal

Lower than 

120

Lower than 

80

Self-check

 

NOTE:

 Show the measured values to your doctor. Never use 

the results of your measurements to change the doses of 

drugs prescribed by your doctor.

3. CONTENTS AND DISPLAY INDICATORS

Model PRO-39

Systolic Pressure

Diastolic Pressure

Pulse Rate

Cuff

Start button

LCD Display

Heart beat symbol
Self-adjustment before measurement
Low battery symbol
Pulse Arrhythmia Detection symbol

4. CONTRAINDICATION

Do not use the device if there is a skin surface damage on the wrist.  

It is inappropriate for people with serious arrhythmia to use this 

device.

5. PRECAUTIONS

1. Read all of the information in the instruction for use before 

operating the unit.

2. Stay still, calm and rest for 5 minutes before blood pressure 

measurement.

3. The cuff should be placed at the same level as your heart.

4. During measurement, neither speak nor move your body and 

arm.

5. Measuring on both arms for each measurement. Hereafter 

measurement should be done on the arm where the blood 

pressure is higher.

6. Please always relax about 5 minutes between measurements 

to allow the blood circulation in your arm to recover. Prolonged 

over-inflation (cuff pressure exceed 300 mmHg or maintained 

above15 mmHg for longer than 3 minutes) of the bladder may 

cause ecchymoma of your arm.

7. Consult your physician if you have any doubt about below cases:

1) The application of the cuff over a wound or inflammation 

diseases;

2) The application of the cuff on any limb where intravascular 

access or therapy, or an arterio-venous (A-V) shunt, is present;

3) The application of the cuff on the arm on the side of a 

mastectomy;

4) Simultaneously used with other monitoring medical 

equipments on the same limb;

5) The usage by a person with a cardiac pacemaker. The device 

does not affect the cardiac pacemaker. However, if there is a 

serious arrhythmia or a low pulse, the measurement results 

may be inaccurate.

8.   This digital automatic blood pressure monitor is designed for 

adults and should never be used on infants or young children. 

Consult your physician or other health care professionals before 

use on older children.

9. Do not use this unit in a moving vehicle.

10. Blood pressure measurements determined by this monitor 

are equivalent to those obtained by a trained observer using 

the cuff/stethoscope auscultation method, within the limits 

prescribed by the American National Standard Institute, 

Electronic or automated sphygmomanometers.

11. Information regarding potential electromagnetic or other 

interference between the blood pressure monitor and 

other devices together with advice regarding avoidance 

of such interference please see part ELECTROMAGNETIC 

COMPATIBILITY INFORMATION.

12. Please do not use the cuff other than supplied by the 

manufacturer, otherwise it may cause biocompatible hazard and 

might result in measurement error.

13.   The monitor might not meet its performance specifications or 

cause safety hazard if stored or used outside the specified 

temperature and humidity ranges in specifications.

14.   Please do not share the cuff with other infective person to 

avoid cross-infection.

15. Please note that changes or modifications not expressly 

approved by the party responsible for compliance could void the 

user’s authority to operate the equipment.

16. This equipment has been tested and found to comply 

with the limits for a Class B digital device, pursuant to part 

15 of the FCC Rules. These limits are designed to provide 

reasonable protection against harmful interference in a 

residential installation. This equipment generates, uses and 

can radiate radio frequency energy and, if not installed and 

used in accordance with the instructions, may cause harmful 

interference to radio communications. However, there is 

no guarantee that interference will not occur in a particular 

installation. If this equipment does cause harmful interference 

to radio or television reception, which can be determined by 

turning the equipment off and on, the user is encouraged to 

try to correct the interference by one or more of the following 

measures:

— Reorient or relocate the receiving antenna.

— Increase the separation between the equipment and receiver.

— Connect the equipment into an outlet on a circuit different 

from that to which the receiver is connected.

— Consult the dealer or an experienced radio/TV technician for 

help.

6. SETUP AND OPERATING PROCEDURES

6.1. Battery loading

a. Open battery cover at the back of the monitor.

b. Load two “AAA” size batteries. Please pay attention to polarity.

c. Close the battery cover.

d. Once you install the batteries or turn off the monitor, the LCD 

does not display anything. Now the monitor is in Off.

 If the LCD displays low battery symbol  , the batteries are 

running out and should be replaced with new ones.

 Rechargeable batteries are not suitable for this monitor.

 Remove the batteries if the monitor will not be used for a 

month or more to avoid relevant damage of battery leakage.

 Avoid the battery fluid to get in your eyes. If it should get in 

your eyes, immediately rinse with plenty of clean water and 

contact a physician.

The monitor, the batteries and the cuff, must be disposed 

of according to local regulations at the end of their usage.

6.2. Applying the cuff

a. Place the cuff around a bare wrist 1-2cm 

above the wrist joint on the palm side of 

the wrist.

b. Place the arm with the cuffed wrist in 

front of your body on a desk or table with 

the palm up. If the cuff is correctly placed, 

you can read the LCD display.

c. The cuff must be neither too tight nor too 

loose. At the same time, there must be 

no free space between the cuff and the 

wrist.

6.3. Carrying out a measurement

Before the measurement:

 Avoid eating, smoking as well as all forms of exertion directly 

before the measurement. All these factors influence the 

measurement result. Try and find time to relax by sitting in an 

armchair in a quiet atmosphere for about five to ten minutes 

before the measurement.

 Measure always on the same arm.
 Attempt to carry out the measurements regularly at the same 

time of day, since the blood pressure changes during the 

course of the day.

Body posture during measurement

Be seated with your feet flat on the floor, and 

don’t cross your legs.

The first way to measure

a. Place palm upside in front of you on a flat 

surface such as a desk or table.

b. Place something under your arm (for example, 

the monitor storage bag) so that the middle of 

the cuff was at the level of the heart. Make sure that the cuff is 

not pressed by anything.

The second way to measure

a. Use the free hand to take the elbow of the arm with the blood 

pressure monitor.

b. Place the arm with the monitor so that the palm of the arm 

would be next to the opposite shoulder, and the monitor - at the 

level of the heart.

c. Make sure that you see the display of the monitor. Relax your 

wrist joint and wrist (do not move your wrist forward or back, 

do not clench into a fist).

Common sources of error:

 Movement during measurement

 The monitor is not at the level of the heart

 The cuff does not fit you in size

 Loose cuff

6.4. Taking your blood pressure reading

a. After applying the cuff and your body is in a comfortable 

position, press the “START” button. All display characters are 

shown for self-test (Picture 1). Please contact the service center 

if any segment is missing on the display.

b. The LCD will momentarily display the result of last measurement. 

See picture 1-1. After the new power the LCD will momentarily 

display 0.See picture 1-2.

         Picture 1              Picture 1-1            Picture 1-2            Picture 1-3

c. Then the monitor inflates the cuff until sufficient pressure has 

built up for a measurement. Then the monitor slowly releases 

air from the cuff and carries out the measurement. Finally the 

blood pressure and pulse rate will be calculated and displayed 

on the LCD screen. Irregular heartbeat symbol (if any) will blink 

on the screen. See picture 1-3. 

d. After measurement, the monitor will turn off automatically 

after 1 minute of no operation. Alternatively, you can press the 

“START” button to turn off the monitor manually.

e. During measurement, you can press the “START” button to turn 

off the monitor manually.

 NOTE:

 The monitor can memorize the last result. If you change 

the batteries, the last result will not be saved.

 NOTE:

 Please consult a health care professional for 

interpretation of pressure measurements.

6.5. Pulse Arrhythmia Detection
Appearance of the Pulse Arrhythmia Detection

Symbol     means that a certain pulse irregularity was detected 

during the measurement. In this case, the result can vary from your 

actual blood pressure that’s why you should rest for 15 minutes and 

repeat the measurement.

The appearance of symbol     is accompanied by a sound signal. 

As a rule this is not a cause for concern; however, if the symbol 

appears more frequently, we recommend you inform your doctor.

The monitor does not replace the cardiological examination, but it 

allows you to detect the arrhythmia even in early stages.

6.6. Technical alarm description

The monitor will show ‘HI’ or ‘Lo’ as technical alarm on LCD with 

no delay if the determined blood pressure (systolic or diastolic) is 

outside the rated range specified in part SPECIFICACIONS. In this 

case, you should consult a physician or check if your operation 

violated the instructions.

The technical alarm condition (outside the rated range) is preset 

in the factory and cannot be adjusted or inactivated. This alarm 

condition is assigned as low priority according to IEC 60601-1-8.

The technical alarm is non-latching and need no reset. The signal 

displayed on LCD will disappear automatically after about 8 

seconds.

6.7. Troubleshooting (1)

PROBLEM

POSSIBLE CAUSE

SOLUTION

LCD Display 

shows 

abnormal 

result

The cuff position was 

not correct or it was not 

properly tightened

Apply the cuff correctly 

and try again

Body posture was not 

correct during testing

Review the “BODY 

POSTURE DURING 

MEASUREMENT” 

sections of the 

instructions and re-test

Speaking, arm or body 

movement, angry, 

excited or nervous 

during testing

Re-test when calm and 

without speaking or 

moving during the test

Irregular heartbeat 

(arrhythmia)

It is inappropriate 

for people with 

serious arrhythmia 

to use this Electronic 

Sphygmomanometer

1-2 cm

IEC 80601-2-30 : 2009+Cor.2010 (Medical electrical equipment – 

Part 2-30: Particular requirements for the basic safety and

essential performance of automated non-invasive 

sphygmomanometers),

EN 1060-1: 1995 + A1: 2002 + A2: 2009 (Non-invasive 

sphygmomanometers – Part 1: General requirements),

EN 1060-3: 1997 + A1: 2005 + A2: 2009 (Non-invasive 

sphygmomanometers – Part 3: Supplementary requirements for

electro-mechanical blood pressure measuring systems).

10. SYMBOL INFORMATION

THE OPERATION GUIDE MUST BE READ 

(The sign background colour: blue.The sign 

graphical symbol: white)

WARNING

TYPE BF APPLIED PARTS

(The cuff is type BF applied part)

ENVIRONMENT PROTECTION –

Waste electrical products should not 

be disposed of with household waste. 

Please recycle where facilities exist. Check 

with your local Authority or retailer for 

recycling advice.

MANUFACTURER’S NAME

ARTICLE NUMBER 

SERIAL NUMBER

OPERATING CONDITION,

TEMPERATURE  10˚C ~ 40˚C

STORAGE CONDITION,

TEMPERATURE  –20˚C ~ 50˚C

KEEP DRY

HOUSING INGRESS PROTECTION RATE: 

IP 22 (Protected from the penetration of 

solid bodies with dimension greater than 

12.5 mm. Protected from the penetration 

of vertically falling water drops)

40

10

-20

50

11. WARRANTY INFORMATION

Warranty period is 3 years from the date of purchase for monitor 

and 1 year for cuff. This warranty doesn’t cover any damages 

caused by improper using, and also battery, and packaging. When a 

manufacturing defect is revealed during the warranty period a faulty 

unit would be repaired or, if repairing is impossible, replaced with 

another one.

Manufacturing date is in a serial number: WWYYXXXXX.
The manufacturer may change units partially or completely if 

necessary, without prior notice.

12. ELECTROMAGNETIC COMPATIBILITY 

INFORMATION

For all ME EQUIPMENT and ME SYSTEMS

Table 1 – Emission

Phenomenon

Compliance

Electromagnetic 

environment

RF emissions

CISPR 11

Group 1, Class B

Home healthcare 

environment

Harmonic distortion

IEC 61000-3-2

Class A

Home healthcare 

environment

Voltage fluctuations 

and flicker

IEC 61000-3-3

Compliance

Home healthcare 

environment

 

Table 2 – Enclosure Port

Phenomenon

Basic EMC 

standard

Immunity test levels

Home Healthcare 

Environment

Electrostatic 

Discharge

IEC 61000-4-2

±8 kV contact

±2 kV, ±4 kV, ±8 kV, ±15 

kV air

Radiated RF EM 

field

IEC 61000-4-3

10 V/m

80 MHz – 2.7 GHz

80% AM at 1k Hz

Proximity fields 

from RF wireless 

communications 

equipment

IEC 61000-4-3

Refer to table 3

Rated power 

frequency magnetic 

fields 

IEC 61000-4-8

30 A/m

50 Hz 

 

Table 3 – Proximity fields from RF wireless communications equipment

Test 

frequency

(MHz)

Band (MHz)

Immunity test levels

Professional healthcare facility 

environment

385

380-390

Pulse modulation 18 Hz, 27 V/m

450

430-470

FM, ±5 kHz deviation, 1 kHz sine, 28 

V/m

710

704-787

Pulse modulation 217 Hz, 9 V/m

745
780
810

800-960

Pulse modulation 18 Hz, 28 V/m

870
930

1720

1700-1990

Pulse modulation 217 Hz, 28 V/m

1845
1970
2450

2400-2570 Pulse modulation 217 Hz, 28 V/m

5240

5100-5800 Pulse modulation 217 Hz, 28 V/m

5500

5785

6.8. Troubleshooting (2)

PROBLEM

POSSIBLE CAUSE

SOLUTION

LCD shows low 

battery symbol    

Low Battery

Change the 

batteries

LCD shows “Er 0”

Pressure system 

is unstable before 

measurement

Don’t move and try 

again

LCD shows “Er 1”

Fail to detect systolic 

pressure

LCD shows “Er 2”

Fail to detect diastolic 

pressure

LCD shows “Er 3”

Pneumatic system 

blocked or cuff is too 

tight during inflation

Apply the cuff 

correctly and try 

again

LCD shows “Er 4”

Pneumatic system 

leakage or cuff is too 

loose during inflation

LCD shows “Er 5”

Cuff pressure above 300 

mmHg

Measure again 

after five minutes. 

If the monitor is still 

abnormal, please 

contact the local 

distributor or the 

factory

LCD shows “Er 6”

More than 3 minutes 

with cuff pressure 

above 15 mmHg

LCD shows “Er 7”

EEPROM accessing 

error

LCD shows “Er 8”

Device parameter 

checking error

LCD shows “Er A”

Pressure sensor 

parameter error

No response when 

you press button or 

load battery

Incorrect operation or 

strong electromagnetic 

interference

Take out batteries 

for five minutes, 

and then reinstall all 

batteries

7. MAINTENANCE

1.   Do not drop this monitor or subject it to strong impact.

2.   Avoid high temperature and solarization. Do not immerse the 

monitor in water as this will result in damage to the monitor.

3. If this monitor is stored near freezing, allow it to acclimate to 

room temperature before use.

4.   Do not attempt to disassemble this monitor and do not 

disconnect the cuff from the monitor.

5. If you do not use the moitor for a long time, please remove the 

batteries.

6. It is recommended the performance should be checked every 2 

years or after it has been dropped. Please contact the service 

center.

7. Clean the monitor with a dry, soft cloth or a soft cloth squeezed 

well after moistened with water or diluted detergent.

8. Keep the cuff clean. It is recommended the cuff should be 

disinfected 2 times every week if needed (For example, in 

hospital or in clinic). Wipe the inner side (the side contacts skin) 

of the cuff by a soft cloth squeezed after moistened with a 3 % 

solution of hydrogen peroxide, then dry the cuff by airing.

9. No component can be maintained by user in the monitor. 

10. The external surfaces of the blood pressure monitors and the 

cuff are resistant to disinfection with a 3% solution of hydrogen 

peroxide.

8. SPECIFICATIONS

1. Product name: Wrist Blood Pressure Monitor, model: PRO-39

2. Classification: Internally powered, Type BF applied part, IP22,

No AP or APG, Continuous operation

3. Machine size: Approx. 73 mm×62 mm×28,5 mm

(2 7/8”x 2 1/2”x 1 1/8”)

4. Wrist cuff circumference: 14cm-19.5cm(approx.5 1/2”-7 11/16”)

5. Weight: not more than 69,3 g ± 10% (exclude batteries)

6. Measuring method: Oscillometric method, automatic inflation 

and measurement

7. Memory Capacity: The last measurement is saved.

8. Power source: batteries: 2 ×1.5V     SIZE AAA

9. Measurement range: Cuff pressure: 0-300 mmHg,

systolic: 60-280 mmHg, diastolic: 20-199 mmHg,

pulse rate: 40-200 beats/minute.

10. Accuracy: pressure:  ±3mmHg, pulse rate:  ±5%.

11. Environmental temperature for operation: 10°C~40°C 

(50°F~104°F).

12. Environmental humidity for operation: ≤85% RH.

13. Environmental temperature for storage and transport: 

-20°C~70°C (-4°F~122°F).

14. Environmental humidity for storage and transport: ≤85% RH.

15. Environmental pressure: 80kPa-105kPa.

16. Battery life: Approx 270 times.

17. Blood pressure monitor set: blood pressure monitor – 1 piece, 

wrist cuff – 1 piece, bladder – 1 piece, AAA batteries – 2 pieces, 

the instruction manual – 1 piece, box – 1 piece, a storage bag –

1 piece.

 NOTE:

 These specifications are subject to change without 

notice.

9. APPLIED STANDARDS

The digital automatic blood pressure monitor corresponds to the 

below standards:

IEC 60601-1:2005/EN 60601-1:2006/AC:2010 (Medical electrical 

equipment – Part 1: General requirements for basic safety and 

essential performance),

IEC60601-1-2:2007/EN 60601-1-2:2007 /AC:2010 (Medical 

electrical equipment – Part 1-2: General requirements for 

basic safety and essential performance – Collateral standard: 

Electromagnetic compatibility – Requirements and tests),

IM_PRO39_EN_PL_425x320_3917

Caring for everyone

Caring for everyone

 

B.Well Swiss AG,

Bahnhofstrasse 24,

9443 Widnau, Switzerland
www.bwell-swiss.ch

Отзывы: