128
T405S Indicator Service Manual
11.2.1 Additional token tables
DIS token table: AWTX Status Byte format
Byte 1
Byte 2
Byte 3
Bit 0
1 = Under or OverLoad
1 = Center Zero
0
Bit 1
0
1=Gross
1
Bit 2
1=kg
1=Net
1=Overload
Bit 3
1=lb
0
1=Motion
Bit 4
1
1
1
Bit 5
1
1
1
Bit 6
0
0
0
Bit 7
0
0
0
WSTAT token table
"O"
Overload
"U"
Underload
"M"
Motion
"S"
Stable
"E"
Cal error
TSAT token table
"NT"
no tare active
"T"
pushbutton tare
active
"PT"
preset or enter
tare active
Example
: Stable and valid gross weight in lb unit of measure would return "822"
The WSTAT and TSAT token default length is 6 characters.
Bit
Byte 1
Byte 2
Byte 3
1
1 = Scale in motion
0 = Stable
1 = Underloaded
0 = Not Underloaded
00 = Low Range
01 = Undefined
10 = Undefined
11 = High Range
2
1 = Scale at Zero
0 =Not at Zero
1 = Overloaded
0 = Not Overloaded
3
1 = RAM error
0 = RAM OK
1 = ROM error
0 = ROM OK
1 = Net Weight
0 = Gross Weight
4
Always 0
1 = Faulty Calibration
0 = Calibration OK
1 = Initial Zero Error
0 = Initial Zero OK
5
Always 1
Always 1
Always 1
6
Always 1
Always 1
Always 1
7
Always 0
1 = Byte follows
0 = Last Byte
1 = Byte follows
0 = Last Byte
8
Parity
Parity
Parity
NCI Status Byte Table
Содержание T405S
Страница 1: ...AWT35 501638 Issue AA T405S Weight Indicator Service Manual...
Страница 2: ...BTEK_T405S_s_en_501638 book...
Страница 8: ...8 T405S Indicator Service Manual...
Страница 134: ...134 T405S Indicator Service Manual...
Страница 147: ...T405S Indicator Service Manual 147 13 7 Outline dimensions...
Страница 150: ...150 T405S Indicator Service Manual...
Страница 154: ...154 T405S Indicator Service Manual...
Страница 155: ......
Страница 156: ...B TEK Scale LLC 1510 Metric Ave SW Canton OH 44706 3088...