16
4.3 Important Advice for Storage
CAUTION
• Before being loaded, the device must have reached the temperature set-point.
• The loading should only be stored in the provided baskets and should not exceed the the storage areas. The
total amount of stored blood bags should be evenly distributed in the 4 baskets (max. 12 blood bags per
basket) (image 4). Do not overload the unit.
• Different blood or platelet types should be sorted and stored in a mannner for the user to identify these with
ease.
• The loading must never obstruct the air-inlet-openings of the cabinet.
• Make sure that the stored load does not tilt to the backside.
• Do not store warm items in the devices. They are not designed for fast cooling of warm items.
• Avoid long and / or frequent door openings to prevent that the inner temperature rises to high.
• Make sure that the door is closed securely and that the door gasket seals adequately against the
cabinet.
• The unit is designed as storage for blood bags and platelets. It is not designed for cooling down the stored
product.
• The maximum storage capacity of the MT100 is 21,5L of blood (48 blood bags of 500ml filled with 450ml
each).
• Store the blood bags in an upright positions (image 3).
image 3: Upright position of blood bags
image 4: Max. storagecapacity
Содержание MT100
Страница 1: ...820 9506 71 ed0821 INSTRUCTIONS FOR USE MT100 BLOOD REFRIGERATOR INCUBATOR...
Страница 2: ...2...
Страница 5: ...5 1 Model Overview Group Model name Blood Refrigerator Incubator MT100...
Страница 24: ...24 12 Spare parts Container door and inner equipment...
Страница 25: ...25 Electronic and cooling unit...
Страница 27: ...27 13 Wiring diagram 294 8062 37...
Страница 28: ...B Medical Systems S r l op der Hei 17 L 9809 Hosingen Luxembourg...