Cardiac Pro Package 287
To adjust ECG offset:
•
Click Offset on the right side of the screen and drag to select the required value.
The available values range from 0 (nearest to the top of the image) to 4 (nearest
to the bottom of the image).
Heart Rate
You can control the number of cycles over which HR is averaged.
To adjust HR average length:
•
Click Heart Rate and drag to select the required number of cycles.
Refractory Time
Refractory time sets the period following the R-peak in which no new R-peak is
acknowledged. This can be used to discriminate the R-peak from the S-peak in
difficult situations.
To adjust refractory time:
•
Click Hold on the right side of the screen and drag to the required value.
The available values range from 0 to 2000 ms in steps of 100ms.
Reviewing ECG
The ECG curve can be stored and reviewed with a cine recording. For information
about using Cine see See “Cine” on page 80.
Содержание Pro Focus 2202
Страница 1: ...English BB1279 A June 2005 Pro Focus 2202 Extended User Guide ...
Страница 14: ...14 ...
Страница 15: ...Part 1 Basics ...
Страница 16: ......
Страница 32: ...32 Chapter 1 ...
Страница 48: ...48 Chapter 2 ...
Страница 49: ...Part 2 Working with the Image ...
Страница 50: ......
Страница 98: ...98 Chapter 5 ...
Страница 117: ...Part 3 Imaging Modes ...
Страница 118: ......
Страница 136: ...136 Chapter 8 ...
Страница 152: ...152 Chapter 10 ...
Страница 164: ...164 Chapter 12 ...
Страница 165: ...Part 4 Setting up and Maintaining Your System ...
Страница 166: ......
Страница 200: ...200 Chapter 13 ...
Страница 208: ...208 Chapter 14 ...
Страница 209: ...Part 5 Pro Packages ...
Страница 210: ......
Страница 288: ...288 Chapter 19 ...
Страница 313: ...Part 6 Appendixes ...
Страница 314: ......
Страница 344: ...344 Appendix C ...