42
Start-up and trumpet curves
15.3
Alarm times
The following graphics show alarm times
depending on pressure level.
15.3.1
Infusomat® Plus Line
00:00:06
00:01:09
00:11:17
00:34:12
02:08:02
08:34:26
00:00:01
00:00:09
00:01:26
00:14:24
02:24:00
14:00:00
P1 =
67 mBar
P9 =
1000 mBar
0,1 ml/h
1 ml/h
25 ml/h
Zeit
(hh:mm:ss, logarithmisch)
Infusomat® Plus Line PUR
Druckstufe
Time [h:mm:ss]
Pressure level
0.1 ml/h
1 ml/h
25 ml/h
P1 =
67 mBar
P9 =
1000 mBar
00:00:05
00:01:07
00:08:48
00:44:15
02:25:55
12:44:23
00:00:01
00:00:09
00:01:26
00:14:24
02:24:00
14:00:00
0,1 ml/h
1 ml/h
25 ml/h
Zeit
(hh:mm:ss, logarithmisch)
Infusomat® Plus Line PVC
P1 =
67 mBar
P9 =
1000 mBar
Druckstufe
Time [h:mm:ss]
Pressure level
0.1 ml/h
1 ml/h
25 ml/h
P1 =
67 mBar
P9 =
1000 mBar
These graphs show the accuracy and
uniformity of flow over time. Take into
account:
Q
The delivery behaviour and the delivery
accuracy are fundamentally affected
by the disposable item used.
Note:
The system accuracy is normally ±5%
of the volume, measured using the trumpet
curve test method according to IEC 60601-
2-24 at a rate of 1 ml/h (at 20 °C ± 2 °C)
and using the recommended disposable
item.
Trumpet curves
(Measured values for second hour in
each case)
Measurement
interval
Δ
t = 0.5 min
Observation
interval
p x
Δ
t [min]
Start-up curves
Measurement
interval
Δ
t = 0.5 min
Measurement
duration
T = 120 min
Flow Q
i
(ml/h)