33
Operation
Q
Remove the infusion line.
Q
Close the pump door while simultane-
ously pulling the door lever.
8.14
Switching off the device
Q
Infusion ended
Note:
The device cannot be switched off if
a disposable item is inserted. Instead it will
go into standby mode.
Q
Press the
On/Off key
for approx.
1.5 seconds.
The device switches off.
8.11
Resetting the therapy
The “Reset therapy” function is used to
delete all currently set therapy data. A new
therapy can be started.
Note:
“
Reset therapy
” can only be selected
if the therapy has been stopped.
Q
Press the
menu key
and select
Reset
therapy
and press
OK
to confirm.
Q
Press the
up arrow key
to reset the
therapy.
Note:
the infused volume (inf. vol.) is not
reset.
8.12
Changing the infusion line
Q
Press the
Start/Stop key
to stop the
infusion.
The green LED turns off.
Q
Disconnect the line from the patient
and close the roller clamp.
Q
Open the pump door by pulling the
door lever. Grip the door lever from
behind and pull it forward.
Q
Remove the infusion line.
Q
Insert the new infusion line,
see section 8.2.
Q
Start the infusion, see section 8.4.
8.13
Ending the infusion
Q
Press the
Start/Stop key
to end the
infusion.
The green LED turns off.
Q
Disconnect the line from the patient
and close the roller clamp.
Q
Open the pump door by pulling the
door lever. Grip the door lever from
behind and pull it forward.