29
Operation
8.8
Using the drug database
DANGER! Danger to the patient from
incorrectly selected drug.
Q
Ensure that the correct drug has
been selected.
Up to 3,000 freely selectable drug names,
including corresponding therapy data and
information and up to 10 concentrations
per drug in 30 categories, can be stored.
The data are loaded using a separate PC
programme.
The drug database can be used to select a
drug name with saved therapy data.
The procedure for selecting a drug is
described below:
Q
Pump has just been switched on or
“Reset therapy” has been selected.
Q
Press the
Menu key.
The main menu is displayed.
Q
Select
Drug
and press
OK
to confirm.
Q
If there is more than one profile avail-
able:
–
Select station
and press
OK
to
confirm.
–
Select patient profile
and press
OK
to confirm.
Q
Select drug category
and press
OK
to
confirm.
Q
Select drug
and press
OK
to confirm.
Q
If available, read the information in
the “Drug info” screen and press
OK
to
confirm.
Q
If necessary,
select concentration
and
press
OK
to confirm.
Q
Read the information in the “Drug”
screen and press
OK
to confirm.
Q
Enter the delivery rate
.
Note:
Manual bolus administration is
limited to a maximum of 10 s. The bolus
administration is automatically stopped,
but it can be continued by pressing the
Bolus key again.
Note:
An acoustic signal sounds for every
1 ml of bolus volume delivered
8.7.2
Administering a bolus with
preselected bolus volume/bolus
duration
WARNING! Danger to the patient
from an overdose. At a bolus rate of
1,200 ml/h, 1 ml is reached after 3 s.
Q
Press the
OK key
to interrupt bolus
administration.
Q
Press the
Bolus key
to call up the
bolus menu.
Entering the bolus volume
Q
Press the
left arrow key
and enter the
desired
bolus volume.
Q
Press the
Bolus key
to start bolus
administration.
Entering the bolus duration (optional)
Q
Press
OK
to confirm the entry of the
bolus volume.
Q
Select
Bolus duration
and press
OK
to confirm.
Q
Entering the desired
bolus duration
.
The bolus rate is calculated.
Q
Press the
Bolus key
.
The bolus administration is started.
After the time has elapsed, the bolus
administration is ended and the infu-
sion continued.