28
Operation
8.7
Administering a bolus
There are three different options for bolus
administration:
Q
Manual bolus
Q
Bolus with preselection of the bolus
volume
Q
Bolus with preselection of the bolus
volume and the bolus duration
Note:
If the bolus administration is not
started after the Bolus key is pressed, the
device automatically returns to the delivery
screen for the ongoing infusion.
Note:
The pressure alarm threshold is
automatically increased to 9 during bolus
administration.
8.7.1
Administering a manual bolus
Q
Press the
Bolus key
.
The “Bolus” screen is displayed.
Q
Press the
Bolus key
again and hold it
down.
Fluid is delivered as long as the key is
pressed or until the maximum duration/
dose have been reached. The delivered
bolus volume is displayed.
Q
Release the
Bolus key
.
The bolus administration is ended and
the infusion continued.
8.6
Activating standby
In the event of longer interruptions, the user
has the option of retaining the set values
and continuing the infusion at a later time.
Activating standby mode
Q
Infusion line inserted and selected
Q
Press the
On/Off key
and hold it down
until the pump indicates that it is in
standby mode.
Adjusting device standby time
Q
Press the left arrow key
.
Q
Enter the desired
time
and press
OK
to confirm.
Ending standby mode
Q
Press
the On/Off key
or
Back key.
Q
Press the
Start/Stop key
.
The delivery is re-started with the
previously set values.