25
Operation
–
Push the infusion line into the air
sensor on the left and into the
pump based safety clamp on the
right.
Ensure that both clips (orange and
white) are firmly latched.
Q
Close the pump door while simultane-
ously pulling the door lever.
Do not damage the infusion line.
8 Operation
Q
Device settings configured
8.1
Switching on the device
Q
Device connected to the mains elec-
tricity or battery fully charged.
Q
Press the
On/Off key
on the device.
The device will perform a self-test:
Note:
Pay attention to audible and visible
alarms, the lighting up of the two status
LEDs and the display during the self-test.
Note:
Alternatively, the device can be
switched on by opening the door.
8.2
Inserting the infusion line
Q
Device switched on.
Q
Open the pump door by pressing the
door lever. To do this, grip the door
lever from behind and pull it forward.
CAUTION! Free-flow
Q
Make sure that the roller clamp is
closed before inserting the line.
Q
Never leave the pump unattended
while the line is being inserted.
Q
Always insert the line completely
and then close the door.
Q
Inserting the infusion line:
–
Insert the green safety clamp into
the green socket on the pump,
ensuring that the push button grips
it firmly.
–
Clip the orange clip into the orange
socket of the pump. The pump is
now holding the infusion line.