7
Validated cleaning and disinfection procedure
Validated procedure
Specific requirements
Reference
Manual cleaning with
immersion disinfection
■
Cleaning brush: 30 mm/
∅
: 4.5 mm, e.g. TA no. 011944
■
20 ml disposable syringe
■
When cleaning products with
movable hinges, ensure that these
are in an open position and, if
applicable, move the joint while
cleaning.
■
Drying phase: Use a lint-free cloth
or medical compressed air
Chapter Manual cleaning/disinfection and
sub-chapter:
■
Chapter Manual cleaning with immer-
sion disinfection
Manual pre-cleaning
with brush and subse-
quent mechanical alka-
line cleaning and ther-
mal disinfection
■
Cleaning brush: 30 mm/
∅
: 4.5 mm, e.g. TA no. 011944
■
20 ml disposable syringe
■
Place the instrument in a tray
that is suitable for cleaning
(avoiding rinsing blind spots).
■
Connect components with lumens
and channels directly to the rins-
ing port of the injector carriage.
■
Keep working ends open for
cleaning.
■
Place instruments in the tray with
their hinges open.
Chapter Mechanical cleaning/disinfection
with manual pre-cleaning and sub-chap-
ter:
■
Chapter Manual pre-cleaning with a
brush
■
Chapter Mechanical alkaline cleaning
and thermal disinfecting