Unit Elements
4
Infusomat® fmS 2.3 gb
4 - 19
4.13
Barcode Label
Designation
Ord. No.
Barcode label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 9070
Exchange
1.
Remove old barcode label, if existing.
2.
Clean adhesion surface with an alcoholic cleaning agent and
let dry.
3.
Loosen barcode label from the base material and stick it on.
Note
Destroy the type plate delivered.
Check
Check that serial number and pump symbol in the plain text field
of the barcode label correspond with the type plate on the pump
of the Infusomat® fmS.
4.14
Frame with Seal
Designation
Ord. No.
Seal plate between frame and front frame
(exchange not recommended) . . . . . . . . . . . . . . . . 3477 3240
Frame with seal plate . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1762
Exchange
1.
Disassemble all parts as described before and exchange frame
with seal plate.
Check
Electrical safety, functional check, pump unit check.
Fig.: 4 - 18
Barcode label
Содержание 871 5416
Страница 1: ...Service Manual Infusomat fmS 0 Version 2 3 English 0 10 ...
Страница 10: ...Contact Persons 0 0 10 Infusomat fmS 2 3 gb For your notes ...
Страница 22: ...System Overview 1 1 12 Infusomat fmS 2 3 gb For your notes ...
Страница 30: ...Software 2 2 8 Infusomat fmS 2 3 gb For your notes ...
Страница 60: ...Unit Elements 4 4 20 Infusomat fmS 2 3 gb For your notes ...
Страница 64: ...Maintenance 6 6 2 Infusomat fmS 2 3 gb For your notes ...
Страница 78: ...Test Equipment and Special Tools 9 9 2 Infusomat fmS 2 3 gb For your notes ...
Страница 82: ...Spare Parts List 10 10 4 Infusomat fmS 2 3 gb For your notes ...
Страница 84: ...Index 11 11 2 Infusomat fmS 2 3 gb ...
Страница 86: ...Revision Documentation A A 2 Infusomat fmS 2 3 gb For your notes ...