Azbil Corporation
Operation
Model SLX 110/120 - Smart displacement type level transmitter 2-75
Operation procedure
Zero adjustment to adjust to real level (pinpoint adjustment)
• Example of use: To adjust output to reading of liquid level of glass gauge
•
Adjustment range: -1.25% (output 3.8 mA) to +105% (20.8 mA)
• Even in the case of engineering unit value (real scale), the digital display in
“pinpoint adjustment” is a PV value display.
Step
Procedure
LCD screen
1
Hold the switch counterclockwise (-) for
approximately 3 sec.
2
When “ZERO DOWN
q
” appears, on the display
turn switch OFF. This means that the “Zero
adjustment mode to adjust to real level” has been
activated.
ZERO DOWN
3-1
As the switch is held counterclockwise (-) the
readout decreases by increments of 0.1%.
When the desired readout appears, turn the switch
OFF.
The normal condition is restored after
approximately 10 sec.
ZERO DOWN
3-2
As the switch is held clockwise (+), the readout
increases by increments of 0.1%.
When the desired readout appears, turn the switch
OFF.
The normal condition is restored after
approximately 10 sec.
ZERO DOWN
Содержание SLX 110
Страница 1: ...Smart Displacement type Level Transmitter Model SLX 110 120 User s Manual CM2 SLX100 2001...
Страница 4: ......
Страница 8: ...iv Model SLX 110 120 Smart displacement type level transmitter...
Страница 40: ...Installation Azbil Corporation 1 26 Model SLX 110 120 Smart displacement type level transmitter...
Страница 120: ...Operation Azbil Corporation 2 80 Model SLX 110 120 Smart displacement type level transmitter...
Страница 150: ...Maintenance Azbil Corporation 3 30 Model SLX 110 120 Smart displacement type level transmitter...
Страница 162: ...Azbil Corporation 5 2 Model SLX 110 120 Smart displacement type level transmitter...
Страница 163: ......
Страница 166: ......
Страница 168: ......