Functions, Configurations, and Structures of model JTD720A and communicator
Azbil Corporation
1-2
Model JTD720A - Smart Multivariable Transmitter
×
Differential pressure DP
Density calculated from static pressure
Density calculated from design pressure
Static pressure PP
DP–LRV
URV–LRV
Figure 1-4 Block diagram of calculation by transmitter with temperature/pressure
compensation (for saturated steam)
LRV: Lower limit of the differential pressure to output 0 % (4 mA DC)
URV: Upper limit of the differential pressure to output 100 % (20 mA DC)
1-2 : Digital indicator (optional)
The display unit of a digital indicator consists of the following:
Main display
Sub display
%
D
H
k g /
a G
-
C
0 0
S T
- 0 . 0 0 M P
1
@
Figure 1-5 Digital indicator
Main display
The main display shows an output of a transmitter in % .
The indications in a measurement mode and in a calibration mode are as follows:
Measurement mode
Compensated output is displayed in % or engineering unit.
Calibration mode
DP mode:
Displays 0 to 100% of setting DP range
PP mode:
Displays 0 to 100% of setting PP range
PT mode:
Displays 0 to 100% of setting PT range
OUTPUT mode
Displays the output set as constant current in %.
At DSP , the value of which it extracts the square root is displayed.
Example: When the output is set to 50%
LIN:
50.00%
DSP
:
70.71%
OUT :
50.00%
Содержание JTD720A
Страница 1: ...CM2 DST720 2001 Smart Multivariable Transmitter Model JTD720A User s Manual...
Страница 34: ...Installation Azbil Corporation 2 16 Model JTD720A Smart Multivariable Transmitter...
Страница 49: ...Azbil Corporation Starting and Stopping this Device Model JTD720A Smart Multivariable Transmitter 3 15...
Страница 50: ...Starting and Stopping this Device Azbil Corporation 3 16 Model JTD720A Smart Multivariable Transmitter...
Страница 70: ...Azbil Corporation 4 20 Model JTD720A Smart Multivariable Transmitter...
Страница 71: ......
Страница 74: ......
Страница 76: ......