10-1
Chapter 10 Self-diagnosis
CAUTION
Do not perform the SFN communication when the process of the liquid level transmitter is in the
automatic control status. Doing so may cause the output to vary, resulting in a hazardous operation
status.
The self-diagnosis results saved into the liquid level transmitter are displayed using the communicator (CommStaff).
Serious Failure (Critical Status)
Diagnostic message
Meaning
Remedy
METER BODY FAULT
AD conversion error
The meterbody may be malfunctioning.
Contact Azbil or your salesperson for assistance.
CHARA PROM FAULT
Character data error
SUSPECT INPUT
Pressure sensor error
MDU/DAC COMP FAULT
CPU error
The electronics module may be malfunctioning.
Contact Azbil or your salesperson for assistance.
RAM FAULT
RAM error
NVM FAULT
NVM error
ROM FAULT
ROM error
PAC FAULT
Output circuit error
INVALID DATABASE
Device setting data error
Minor Failure (Non-Critical Status)
Diagnostic message
Meaning
Remedy
SENSOR OVER TEMP
Pressure sensor temperature
error
Modify the pressure sensor installation so that the
meterbody temperature is within the operating range. Or,
check the process.
EXCESS ZERO CORR
The zero calibration amount
exceeds ±5 % when compared to
the time of shipment.
Check if the input pressure is zero or the span, and then
recalibrate.
EXCESS SPAN CORR
The span calibration amount
exceeds ±5 % when compared to
the time of shipment.
IN OUTPUT MODE
Constant current mode
Cancel the constant current mode.
M.B. OVERLOAD OR
Excessive pressure or meterbody
error
Check that the input pressure is within the specification
range.
If the input pressure is too great, reduce it or change the
device to a model with a larger range.
METER BODY FAULT
Undefined*
--
CORRECTS RESET
Noncalibrated
Calibrate the lower range value and upper range value of
the setting range.
NO DAC TEMP COMP
Undefined*
--
EXT.SWITCH FAULT
Undefined*
--
* This ALTJ9000 model does not use these items.
Содержание ALTJ 9000 Series
Страница 1: ...CM2 JTL100 2001 ALTJ 9000 Immersion Type Liquid Level Transmitter User s Manual...
Страница 14: ...1 6...
Страница 18: ...2 4...
Страница 28: ...6 2...
Страница 34: ...7 6...
Страница 38: ...8 4...
Страница 58: ...9 20...
Страница 60: ...10 2...
Страница 64: ......
Страница 65: ......
Страница 66: ......
Страница 68: ......