Avisoft UltraSoundGate 116Hm Скачать руководство пользователя страница 1

Components of the UltraSoundGate 116Hm

7

12

Introduction

Thank you for purchasing Avisoft UltraSoundGate 116Hm. This 
bus-powered USB device supports single-channel high-speed data 
acquisition at sampling rates of up to 1 MHz.
The accompanying recording software Avisoft-RECORDER USGH 
provides either continuous or triggered direct-to-disk recording with 
real-time spectrogram displays.

Getting started

The supplied RECORDER USGH software can be launched  from 

Start / All Programs

 / 

Avisoft Bioacoustics 

RECORDER USGH 

On the first program start, the configuration dialog box will be 
launched automatically (otherwise it is available from 

Options

 / 

Configuration). 

Select the desired 

Sampling rate

 from the 

Input 

Device Settings 

section and click at 

Ok. 

Then click at the Pause 

button (

Monitoring/Pause

) and the Start button (

Monitoring/Start

). 

You will then see the real-time spectrogram displaying the incoming 
signals. 

11

1 XLR input connector

The 5-pole XLR input connector represents the analog inputs of the 
recording device and provide power supply voltages for external 
amplifiers and microphones. The connector scheme is as follows:

 

2 SYNC input/output

This socket allows to synchronize the sample clock signals of several 
UltraSoundGate units. To accomplish this, all devices must be 
connected to a single computer and one device must be configured 
as the master (see section 

RECORDER USGH Settings

 for details).

3 TRIGGER input/output

This 2-pole (mono) 2.5 mm mini-jack connector is electrically 
connected to the TRG button (7) and allows connecting an external 
trigger. This input is TTL-compatible (there is additionally an internal 
pull-up resistor of 10 kOhm to Vcc). Pulling this input to ground (for 
instance by closing a simple switch) will activate the logic 

USG TRG 

button

See the last page on how to congigure the TRIGGER socket 

as an output

1

2

3

-

+

differential source

4

5

+200V

+5V

1

2

3

-

+

single-ended source

4

5

+200V

+5V

4 DIN

This 2-pole 2.5mm mini-jack connector allows to connect an external 
digital signal. The input is TTL-compatible (internal pull-up resistor of 
10 kOhm to Vcc). The status of this signal is stored in the LSB (bit 0) 
of the 16-bit data words that are transmitted over the USB and can be 
used as a sample-precise trigger source in the RECORDER software. 
It can be extracted afterwards by the Avisoft-SASLab Pro sound 
analysis software (e.g. for creating labels). The digital input 
functionality is not available in the 8-bit recording mode.

 

5 PHONES 

This stereo 3.5mm mini jack connector allows to connect headphones 
or a small speaker for acoustically monitoring the incoming 
ultrasounds. A undersampling technique is being used to convert the 
ultrasonic sounds into audible signals. The undersampling ratio can be 
adjusted from the Advanced USGH Settings dialog box of the 
RECORDER USGH software (see section 

RECORDER USGH 

Settings 

for details).

6 TRIGGER button

This button can control the .wav file recording process in the 
RECORDER software. To enable this mode of operation, one of the 
following 

Trigger

 source options must be selected from the 

configuration dialog box:

USG TRG button auto hold

 : Pressing the button for more than 2 seconds will 

activate an auto hold mechanism (the recording continues after releasing the button and 
will stop once the button is pressed again). If the button is pressed for less than two 
seconds, it will only record as long as the button is being pressed.

USG TRG button

 : The software will record as long as the button is pressed.

USG TRG button inversed

 

:  The software will record as long as the button is not 

pressed (or as long as the external TRG signal is not active (logic high)). 

USG TRG button toggled

 : The software will start recording once the button is 

pressed and continues until the button is pressed again.

7 REC indicator

This amber colored LED will flash once the device is connected to the 
PC. It will be switched off once the RECORDER USGH software is 
running the in the monitoring mode. In this mode, the REC LED 
indicates whether the RECORDER software is recording the incoming 
data onto disk. 

8 Peak level meter

The peak level meter indicates the instantaneous recording level The 
red OVER LED indicates clipping (over-modulation). If this happens, 
the gain should be reduced by turning the gain control knob to the left 
(anticlockwise).

9 POWER indicator

This green LED indicates that the unit is connected to the USB power 
supply.

10 GAIN control knob

This control knob adjusts the analog input recording level.

11 VOLUME control knob

This knob adjusts the volume of the phones monitor output (5).

12 USB 2.0 interface

The supplied USB cable with the push-pull plug must be plugged in 
here. The USB cable should be first connected to the UltraSoundGate 
116Hm and then to the computer. Make sure that the red dots on the 
LEMO socket and the plug are aligned. Otherwise the device can be 
damaged. It is also recommended to leave the USB cable connected 
to the UltraSoundGate when it is not in use.

  

8

9

6

4

5

1

3

2

10

1 Ground
2 Positive input
3 Negative input
4 +5V supply voltage (max current 20 mA)
5 +200V polarization voltage

Installation procedure

First download and install the RECORDER USGH software from the 
Avisoft Bioacoustics website (www.avisoft.com/downloads.htm or 
directly www.avisoft.com/RECORDER USGH.exe). This installation 
program will install both the RECORDER USGH application 
(rec_usgh.exe) and the required device drivers (usgh_xx16h.inf, 
usgh.sys) for the UltraSoundGate xx16H devices. When the 
installation procedure has completed, the UltraSoundGate unit can 
be connected to the computer. The device should then be detected 
as “Avisoft-UltraSoundGate 116H” and the pre-installed driver 
should be finally activated. 

Under some circumstances it might happen that the silent 
installation of the device driver fails. If that happens, navigate to the 

Windows Control Panel

 > 

Hardware and Sound

 > 

Device Manager

 

and right-click at the entry

 Other devices

 > 

Avisoft UltraSoundGate 

116H

 and select the 

Update Driver Software..

. option. Then click at 

Browse my computer for device driver software,

 click at 

Browse 

and 

navigate to the folder

 C:\Program Files (x86)\Avisoft 

Bioacoustics\RECORDER USGH\Drivers

 and finally click at Next. 

The completed device driver installation will then look like this:

To become familiar with the RECORDER USGH 
software use the online help system that can be 
accessed through the drop-down menu 

Help > Help...

the 

Help...

 buttons on the individual dialog boxes or 

the website at 

http://www.avisoft.com/Help/RECORDER/content.htm

 

and the section

 

RECORDER USGH Software 

Settings

 in this guide.

Содержание UltraSoundGate 116Hm

Страница 1: ...o hold mechanism the recording continues after releasing the button and will stop once the button is pressed again If the button is pressed for less than two seconds it will only record as long as the button is being pressed USG TRG button The software will record as long as the button is pressed USG TRG button inversed The software will record as long as the button is not pressed or as long as th...

Страница 2: ...33333 Hz and an undersampling ratio of 10 1 the resulting output sample rate will be 33333 Hz All input signal components above the Nyquist frequency half of the output sample rate 16666 Hz will be folded down to the range between zero and the Nyquist frequency For instance a 30 kHz input signal will appear at the monitor output at a frequency of 3333 Hz The Settings button in the Input Device Set...

Страница 3: ...ENDOR EXCLUDES ALL IMPLIED WARRANTIES INCLUDING BUT NOT LIMITED TO IMPLIED WARRANTIES OF MERCHANTIBILITY AND FITNESS FOR A PARTICULAR PURPOSE 3 The Vendor s entire liability and the Buyer s exclusive remedy shall be at the Vendor s SOLE DISCRETION either 1 return of the device and refund of purchase price or 2 repair or replacement of the device 4 THE VENDOR WILL NOT BE LIABLE FOR ANY SPECIAL INDI...

Страница 4: ...y min trim 3 2dBV 1 dBu 0 69 Vrms Gain adjustment potentiometer 40 dB continuous range Input impedance 50 kOhm Analog input connector female XLR 5 socket Other inputs external trigger TTL compatible digital input output TTL compatible SYNC in out Computer interface USB 2 0 isochronous high speed mode Max supply current drawn from the USB 250 mA Physical dimensions W H D 58 x 22 x 160 mm Weight 200...

Страница 5: ...n trim 1 6 dBV 3 8 dBu 1 2 Vrms Gain adjustment potentiometer 30 dB range with 3 dB increments Input impedance 50 kOhm Analog input connector female XLR 5 socket Other inputs external trigger TTL compatible digital input output TTL compatible SYNC in out Computer interface USB 2 0 isochronous high speed mode Max supply current drawn from the USB 250 mA Physical dimensions W H D 58 x 22 x 160 mm We...

Страница 6: ...tionCD from the GPS subfolder on the Avisoft Bioacoustics CD or from the manufacturer website www navilock com Once the GPS receiver is connected to the PC a new USB device will be detected and its USB Serial Controller driver should be finally activated The sucessfully installed driver is listed on the Ports section of the Windows Device Manager as a Profilic to serial Comm Port Windows automatic...

Страница 7: ... is the same because the Bluetooth communication is estabished through a virtual COM port The proviously installed GPS receiver can then be selected from the Avisoft RECORDER command Options GPS Settings By clicking at the Scan ports button in the GPS receiver communication settings section the proper COM port settings for the GPS receiver should be detected automatically Alternatively you might s...

Страница 8: ...to SMPTE Time Code SMPTE time code signals are provided by a number of SMPTE time code generators such as the model HORITA PG 2100 or TG 50 TG 50P The output signal of these devices is an AC coupled signal that is originally intended for recording onto an analog audio input channel It is therefore required to convert this AC coupled signal to the open collector DIN input of the UltraSoundGate A su...

Страница 9: ...frames Note that the time code track located on the least significant bits of the wav file will be destoyed by applying any filter command unles the option keep DIO state is activated SMPTE and LANC time code can also be synthesized by the Avisoft SASLab Pro software and stored into the LSB of 16 bit wav files File Specials UltraSoundGate DIO Insert time code pulse train or Edit Synthesizer Insert...

Страница 10: ... the notebook computer is fixed reliably to the frame before using it Make sure that the velco tape is attched tightly so that the units cannot slip out It is recommended to use a short USB cable to connect the UltraSoundGat eunit to the laptop netbook PC When using the UltraSoundGate with the carrying frame it is recommended to add the RECORDER USGH software to the Windows Startup folder and to a...

Отзывы: