Chapter 3: System Operation
KXL System Operator’s Manual,
Rev G
3-10
3.9 Preparing the Patient
Ensure that the patient is lying flat or reclined on a patient table or chair. His or her
head should rest in a head rest.
Adjust the table or chair and head rest so that the patient can rest comfortably for the
duration of the treatment without head movement.
Apply a lid speculum and optional drapes using standard clinical technique.
Apply Riboflavin to the area of treatment in accordance with the
Riboflavin
Instructions for Use (IFU).
CAUTION: Riboflavin (vitamin B2) is not a part of the KXL System described in this
manual. For details of component use, please refer to the component’s Directions for Use.
Figure 3-11. Prepare Patient Screen
NOTE: Once the Riboflavin is applied to the eye, Start the induction by pressing the
“Riboflavin Applied: Start Timer” button.