Chapter 3: System Operation
KXL System Operator’s Manual,
Rev G
3-3
3.3 UV Dose
The UV irradiance dose is the product of the intensity of the irradiance and the exposure
time. Both intensity of the irradiance and the exposure time are adjustable and displayed.
The system tracks treatment time, intensity of irradiance and total dose during treatment.
NOTE
- The system's parameters are:
Induction Period:
1 – 30 minutes
Irradiance:
3 – 45 mW/cm
2
Exposure Time:
1 – 30 minutes
The maximum dose is 10 J/cm
2
Please reference Riboflavin Instructions for Use (IFU) for formulation information.
3.4 Preparing the System
Position the KXL System adjacent to the treatment table or chair. Lock the casters to
secure the device’s position.
Make sure the system is turned ON.
Check glass window of beam aperture for dust and dirt. See sections 4.8 and 4.9 for
cleaning instructions.
3.5 Important Steps before Turning on the
System
The user is responsible for assuring that the KXL System is functioning properly and is in
good-working condition before starting a treatment.
To ensure the system is functioning properly, consider the following mandatory points:
o
Inspect the device, accessories, and connecting cables for visible damage.
o
Take local regulations for use of portable electro-optical medical devices into
consideration.
3.6 Powering Up the System
Turn ON the single power switch on the front of the KXL console. This switch turns on
all the system components.
The KXL System begins a power-up sequence, loading the operating system and all
configuration and reference files.
Содержание KXL System
Страница 6: ...KXL System Operator s Manual Rev G vi This page is left blank intentionally...
Страница 36: ...Chapter 3 System Operation KXL System Operator s Manual Rev G 3 22 This page is left blank intentionally...
Страница 42: ...Chapter 5 Equipment Classification KXL System Operator s Manual Rev G 5 2 This page is left intentionally blank...
Страница 46: ...This page is left intentionally blank...