KXL System
Operator’s Manual,
Rev G
3-1
3 System Operation
3.1 Charging the KXL Battery
NOTE: Prior to initial use the internal battery pack of the KXL must be charged
overnight.
If using the internal rechargeable battery, at the end of each business day or when not in
use, the KXL should be connected to grounded supply mains to allow charging of the
battery pack.
The change of the battery is identified by a light located on the column of the KXL, the
following describes the battery charge.
o
Orange – low, charging
o
Yellow – charging
o
Green – fully charged
NOTE: if the battery does not appear to be holding a charge or charging please
contact your local Avedro Service Representative.
NOTE: the battery pack on the system should last for 16 hours during normal operation.
The software system will notify the user when the battery needs to be charged. The KXL
is designed to prohibit treatment if there is not sufficient battery power to perform a
treatment. (See chapter 4 Maintenance for more information on troubleshooting battery
problems.)
3.2 Touchpad/Keyboard Use
The table below identifies and describes important touchpad keys and icons unique to KXL
System operation. Chapter 2 identifies and describes the system’s major components.
Touchpad Key
Icon
Description/Function
Power Off button
(Initial screen)
Turns OFF electric power to
the personal computer.
Start New Treatment
button (Initial screen)
Directs the system to repeat
the entire clinical treatment
protocol so the physician
can treat another patient.
Содержание KXL System
Страница 6: ...KXL System Operator s Manual Rev G vi This page is left blank intentionally...
Страница 36: ...Chapter 3 System Operation KXL System Operator s Manual Rev G 3 22 This page is left blank intentionally...
Страница 42: ...Chapter 5 Equipment Classification KXL System Operator s Manual Rev G 5 2 This page is left intentionally blank...
Страница 46: ...This page is left intentionally blank...