Oxygen sensor
:
Clean with a damp cloth soaked in mild detergent and then wipe off the remaining detergent with a
dry lint free cloth.
◼
Absorber canister assembly :
First flush with water, then soak it in solution mixed by water and detergent 30 minutes, the
suggested temperature is 30-41
℃
, after that treat by clean water, finally wipe by using 70%
medical use alcohol or autoclave under maximum temperature 134
℃
.
◼
Reusable manual bag
:
First flush with water, then soak it in solution mixed by water and detergent 30 minutes, the
suggested temperature is 30-41
℃
, after that treat by clean water, finally wipe by using 70%
medical use alcohol or autoclave under maximum temperature 134
℃
.
◼
CO2 circle absorber:
First flush with water, then soak it in solution mixed by water and detergent 30 minutes, the
suggested temperature is 30-41
℃
, after that treat by clean water, finally wipe by using 70%
medical use alcohol or autoclave under maximum temperature 134
℃
.
◼
Bag Arm:
First flush with water, then soak it in solution mixed by water and detergent 30 minutes, the
suggested temperature is 30-41
℃
, after that treat by clean water, finally wipe by using 70%
medical use alcohol or autoclave under maximum temperature 134
℃
.
9.3.1 Re-install the CO2 Circle Absorber
Make sure that all components of breathing system are fully dry before installing it.
9.3.2 Re-install the bellows assembly
1. Make sure that all components of bellows are fully dry.
2. Install the ring, then push the latch toward the center and attach the locking tabs. Install the
pressure-relief valve.
3. Push the latch toward the center and install the rim. A
“Double Click” should be heard when the
rim is installed. Pull up on the rim to make sure it
’s locked.
4. Install the folding bag and bellows housing.
5.install the plastic cover on the base.
9.3.3 Re-install the Bag Arm
When the bag arm is fully dry, Install the Bag Arm.
9.3.4 Re-install the Airway Pressure Gauge
When the airway pressure gauge is fully dry, Install the Airway pressure gauge.
9.3.5 Re-install the Oxygen Sensor
1. Make sure that all components of oxygen sensor is fully dry.