gas leakage. It’s dangerous to patient if the canister is not installed in a long time.
9.3 Cleaning& disinfection and Re-install the CO2 Circle Absorber
and bellow
Parts marked are autoclavable. Metal and glass parts can be steam autoclaved. Maximum
commended temperature is 134ºC. By using autoclave to solidify bacterioprotein rapidly, quick
and reliable disinfection can be achieved. Suffered from 15 to20 minutes of 1.05 kg/cm
2
steam
pressure and 121ºC temperature, all bacteria and most blood cells are killed.
Such parts are washable by hand. Rinse and dry all parts of the breathing system except the O
2
sensor completely by using mild detergent (pH ranging from 7.0 to 10.5).
Warning
◆
Do not use talc, zinc stearate, calcium carbonate, corn starch or equivalent materials to
prevent tackiness. These materials can go into the patient’s lungs and airways and cause
irritation or injury.
◆
Do not put both of the breathing system and the O
2
sensor in liquid or autoclave them.
◆
Inspect all parts for deterioration. Replace them if necessary.
All parts of the breathing system can be cleaned and disinfected. The cleaning and disinfection
methods are different for different parts.
You need to select the appropriate method to clean and disinfect the parts based on the actual
situations to avoid cross-contamination.
This table is our recommended cleaning and disinfection methods for all parts of the breathing
system.
The following is our recommended disinfection method:
◼
Patient breath circuit and Y piece(reusable)
:
First flush off by water, then soak it in solution mixed by water and detergent 30 minutes, the
suggested temperature is 30-41
℃
, after that treat by clean water, finally wipe by using 70%
medical use alcohol or autoclave under maximum temperature 134
℃
.
◼
Flow sensor
:
First flush off by water, then soak it in solution mixed by water and detergent 30 minutes, the
suggested temperature is 30-41
℃
, after that treat by clean water, finally wipe by using 70%
medical use alcohol.
◼
Bellows assembly:
First flush with water, then soak it in solution mixed by water and detergent 30 minutes, the
suggested temperature is 30-41
℃
, after that treat by clean water, finally wipe by using 70%
medical use alcohol.
◼
Check valve assemblies
:
First flush with water, then soak it in solution mixed by water and detergent 30 minutes, the
suggested temperature is 30-41
℃
, after that treat by clean water, finally wipe by using 70%
medical use alcohol or autoclave under maximum temperature 134
℃
.