Compliance with the condition shall be proven by verification
Minimum resolvable temperature difference (MRTD / NETD): max. 0.1°C
Image homogeneity: max. 0.2°C deviation over the entire image or over the
image section used for measuring the face. Compliance with the condition must
be verified by testing
Position of the face
Detection at a height of 0.75 - 2.2m above the ground
Image layer vertical to the face
Switch on
Operation must be prevented during the warm-up phase
Technical alarm condition of at least low priority must be displayed until the
camera is ready for operation (message on screen)
Spatial resolution: <= 1mm / Pixel
Emissivity: 0.98 must be set
Alarm: Display of a physiological alarm condition of at least medium priority when
the temperature threshold is exceeded (graphical alarm display)
Technical Data
IrCamera
General Features
Model
IRSX-I 336S
IRSX-I 640S
Detector type
Focal Plane Array (FPA), uncooled microbolometer
Spectral range
7.5 to 13 μm
Pixel size
17 x 17 μm
Measuring speed
9 Hz
9 Hz
Measurement
Measuring ranges
High Gain
-40°C to 135°C
NETD
< 30 mK @ F/1.0
Lenses
7,5 mm
Measuring distance < 1m
13 mm
Measuring distance 2m
Measuring distance < 1m
25 mm
Measuring distance 2m
Содержание FebriScan
Страница 1: ...April 28th 2020 Page 1 of 49 AT Automation Technology GmbH FebriScan USER GUIDE...
Страница 10: ...April 28th 2020 Page 10 of 49 Wing screw Central fixing screw...
Страница 12: ...April 28th 2020 Page 12 of 49 320mm...
Страница 42: ...April 28th 2020 Page 42 of 49 Appendix Technichal Documentation Blackbody...
Страница 43: ...April 28th 2020 Page 43 of 49 Control electronics Blackbody...
Страница 44: ...April 28th 2020 Page 44 of 49 IRSX I...