11.Maintenance 

 

The micro motor cable has limited durability and is considered a consumable product.  The durability of the cable is largely dependent 
upon use and treatment, i.e., handling, reprocessing and frequency of use.  

Caution must be exercised when connecting or disconnecting the micro motor and foot control cable from the device to avoid damage 
to the cable. 

Periodically inspect the micro motor cord for damage prior to use. If there is damage noted, contact the Manufacturer.   

LUBRICATING: 

-

 

Do not oil or lubricate the micro motor. 

RUBBER O-RING: 

-

 

If necessary, substitute. Contact the Manufacturer  

1.Do not attempt to disassemble the motor or motor connector. 

2.Do not oil or lubricate the motor. 
3. Do not attach a hand piece to the motor while the motor is running. 

4.Do not bend motor cord sharply. 

-

 

The Micro motor is sensitive to shock. Do not drop or impact micro motor against hard surface 

Failure to comply with any of the above instructions may void your warranty 

 

 

12.TECHNICAL DATA 

 

Manufacturer 

ATS Dental – Pistoia – Italy  

Nominal Power 

100 VA 

Model 

OMEGA IMPLANT 

Frequency 

50/60 Hz 

Dimensions 

185x160x100 

Protection 

Class 1 

Materials 

ABS 

Type 

BF 

Micro motor 

Brushless 40000/30000 

Class MD 

IIa rule 9 in Annex IX 93/42/EEC 

Noise 

< 65 dBA 

Console 

IP21 

Supply Voltage 

24-30 VDC  

Micro motor 

and Pedals 

IP21 

Not suitable for use in presence of flammable anesthetics or oxygen. 
Conditions of use: Temp +18°C/+40°C (+64°F/+140°F) RH < 80% 
Shipping and Storage conditions: Temp +5°C/+65°C (+41°F/+149°F) RH < 20-95% non condensing. 

Life of the device: When used as prescribed = 5 years. No warranty claim may be made in the event that the failure happen sooner 
or later than indicated, because of the frequency of use, sterilization and maintenance. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание AOMG300DL Plus

Страница 1: ...l it REF Model AOMG300DL plus Omega Implant Autoclavable Motor DeLuxe Pedal L E D Light OMG300DL Omega Implant NOT Autoclavable Motor DeLuxe Pedal OMG300S Omega Implant NOT Autoclavable Motor Standard...

Страница 2: ...stry for surgery to expose and dissect oral tissue structures such as the periodontal gap gingival bone the jaw extractions and implantations The product may not be used for a purpose for which it was...

Страница 3: ...a rotary instrument holder onto which a tool drill burs etc is attached The rotation of the micromotor activates the tool Part s applied Rotary instrument holder Tool drill bur etc Utilization The me...

Страница 4: ...ser should nevertheless ensure that any potential electromagnetic interference does not cause an additional risk presence of radiofrequency emitters electronic devices etc Interference may occur when...

Страница 5: ...or DeLuxe Pedal Electronic Control Console Micromotor Assembly NOT Autoclavable Pedal DeLuxe Irrigation Bag Hanger Rod Peristaltic Pump Tubing C Power Supply OMG300S Omega Implant NOT Autoclavable Mot...

Страница 6: ...e supplied non sterile except for tubing 5 The serial number on the Unit s rear panel must be the same as the one on transport documents 6 Allow at least 150 mm free space around the unit for cooling...

Страница 7: ...ust Pump Flow 4 Allows the user to store and quickly access up to 5 different Programs 5 Torque adjust 6 Speed Increase Decrease 7 Handpiece ratio selector 8 Forward Reverse Selector 9 Calibrate handp...

Страница 8: ...oard The central down button allows user to control up to the pre selected maximum speed and switch on the pump if enabled REF OMG300S STANDARD REF OMG300E EASY It allows user to control up to the pre...

Страница 9: ...ar panel OMEGA IMPLANT 6 Only with REF AOMG300DL in case switch on on the rear panel the LED Light In every program Programs 1 to 5 are set the same parameters by the manufacturer Perform torque calib...

Страница 10: ...n liquids and sprays Never immerse the unit or its parts in liquid External surface of the console may be cleaned with a damp cloth or disinfectant wipe The housing is not waterproof Do not use ultras...

Страница 11: ...Do not attach a hand piece to the motor while the motor is running 4 Do not bend motor cord sharply The Micro motor is sensitive to shock Do not drop or impact micro motor against hard surface Failure...

Страница 12: ...cable not connected properly Check to see that the micro motor cable is plugged into the connector on the front right of the central unit disconnect and reconnect Contrangle defective or not properly...

Страница 13: ...he relative humidity should be at least 30 Burst Fast transient EN 61000 4 4 2kV for power supply lines 2kV for power supply lines Quality of the Main power source should be that of a typical commerci...

Страница 14: ...each frequency range b Interference may occur in the proximity of equipment marked with the following symbol Recommended separation distance between portable and mobile RF communications equipment an...

Страница 15: ...USER MANUAL Rev 1 April 2017 Omega Implant Page 15 of 16 ATS DENTAL srl 51030 Serravalle Pistoiese Italy www atsdental it NOTE...

Страница 16: ...been validated by the manufacturer of the device as suitable for preparing the device for reuse The proper implementation of the preparation process as well as achieving the desired result are full r...

Отзывы: