14
Electromagnetic compatibility
The fusionTrack is medical electrical equipment and needs special precautions regarding EMC and needs
to be installed and put into service according to EMC information provided in this document.
EMC tests have not included any AC-adapter with the product.
Applied standards
EN 60601-1-2:2007 +AC:2010
IEC 60601-1-2:2007 (ed3.0)
Medical electrical equipment – Part 1-2: General re-
quirements for basic safety and essential performance
– Collateral standard: Electromagnetic compatibility –
Requirements and tests
EN 60601-1-2:2015
IEC 60601-1-2:2014 (ed4.0)
Medical electrical equipment – Part 1-2: General re-
quirements for basic safety and essential performance
– Collateral standard: Electromagnetic compatibility –
Requirements and tests
Electromagnetic emissions
The fusionTrack is intended for use in the electromagnetic environment specified below. The integrator, the customer
or the user should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
EN 55011:2016
CISPR 11:2015 (ed6.0)
Group 1
The fusionTrack uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby elec-
tronic equipment.
RF emissions
EN 55011:2016
CISPR 11:2015 (ed6.0)
Class B
The fusionTrack is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply net-
work that supplies buildings used for domestic pur-
poses.
Harmonic emissions
EN 61000-3-2:2014
IEC 61000-3-2:2014
Not applicable
Voltage fluctuations / Flicker emissions
EN 61000-3-3:2013
IEC 61000-3-3:2013
Not applicable
Use of the fusionTrack adjacent to, or stacked with other
equipment should be avoided because it could result in im-
proper operation and can lead to wrong measurements to
serious injury including death. If such use is necessary, the
fusionTrack and the other equipment should be observed to
verify that they are operating normally.
79