Figure 12.2: Running the zeroing procedure.
- a deformed pointer or
marker
;
- a decalibrated fusionTrack device;
- dirty optics on the fusionTrack device.
12.2
Running the procedure
The zeroing procedure must be integrated in the workflow at each startup of the application or as soon as
the pointer (whole pointer or any of its
fiducials
in case of passive
markers
) is changed. The operator for
this test is the surgeon. How to proceed must be clearly stated in the user manual.
Any liquid on lenses such as body fluids ejection or conden-
sation could lead to wrong measurements. Lenses have to
be cleaned as soon as they get in contact with liquids. The
zeroing method has to be re-performed after each cleaning.
Some disposable passive
fiducials
cannot be cleaned, as the
reflective coating would be damaged, and may require to be
changed during the operation. The zeroing method has to
be re-performed after changing any
fiducials
.