User manual
Version: v4.5.2
The fusionTrack products are intended for indoor use in professional healthcare facility environments
like physician offices, dental offices, clinics, limited care facilities, freestanding surgical centers, freestanding
birthing centers, multiple treatment facilities, hospitals (emergency rooms, patient rooms, intensive care,
surgery rooms) except for:
- near active HF (high frequency) SURGICAL EQUIPMENT,
- inside RF (radio frequency) shielded room of an ME SYSTEM for magnetic resonance imaging, where
the intensity of EM DISTURBANCES is high
- in ambulances.
The calibration of the device is performed at 20 °C. The best accuracy and precision performances are
obtained when the fusionTrack is operated in a room at 20 °C, once thermal equilibrium is reached. The
fusionTrack should be used in the conditions listed in Table 5.1.
Environmental parameter
Range
Temperature
15 °C to 30 °C
Atmospheric pressure
70 kPa to 106 kPa
Relative humidity
20 % to 80 % non condensing
Altitude
−
450 m to 3000 m
Table 5.1: Recommended environmental conditions.
5.2 Least favorable working conditions
The fusionTrack has only one functioning mode, therefore no
least favorable working conditions
exist.
5.3 Transportation and storage environmental conditions
The integrator of the fusionTrack system must take care of the mechanical stability of the fusionTrack at
all times (when the fusionTrack is in use but also when it is not powered). Effectively, the calibration of the
device may be invalidated if the device suffered from shocks, drops and vibrations. The fusionTrack should
therefore be transported and stored in an adequate transportation / storage package which minimises the
probability of shocks, drops and vibrations and absorbs mechanical shocks and stress as good as possible.
If the fusionTrack device was stored at extreme environmental conditions, a reasonable amount of time must
be waited before using it, so that the operating environmental conditions are reached.
Environmental parameter
Range
Temperature
−
30 °C to 60 °C
Atmospheric pressure
50 kPa to 106 kPa
Relative humidity
10 % to 90 % non-condensing
Table 5.2: Recommended short time (
<
1 week) storage and transportation conditions.
The fusionTrack has been designed to be stored and transported in the conditions listed in Table 5.2
and Table 5.3. Note that during both transport and storage, special handling measures have to be taken:
- keep away from direct sunlight;
Atracsys / 2020-06-17 / 16:48:00