Maintenance and service
6 Maintenance and service
Maintenance, repairs and period tests may only be carried out by persons who have
the appropriate technical knowledge and are familiar with the product. The person in
question must possess the necessary test devices and original spare parts required to
carry out these measures.
ATMOS recommends: Work should be carried out by an authorised ATMOS service
partner. This ensures that the repairs and testing are carried out professionally, that
original spare parts are used and that warranty claims remain unaffected.
Maintenance, repairs and period tests must
not
be carried out while the product is
being used on the patient.
6.1 Period tests
ATMOS® S 61 CORIAN® instruments/ATMOS® S 61 CORIAN® instruments XXL
None
ATMOS® S 61 CORIAN® integral
Observe the specifications for the individual components.
6.2 Function check
Perform a function check at least once a week, and also at any other time if you have
any concerns with regard to the safety of the device.
Do not operate the product if you notice any damage. In this case, clean and disinfect
the device and send it to ATMOS for repair.
All variants
1. Check that the product is not damaged.
2. Check that the product is in good hygienic condition.
3. Check that all the moving parts can be used without any problems:
• Drawer
• Drawer for devices
• Waste disposal
• Instrument disposal
ATMOS® S 61 CORIAN® integral
1.
Check whether the fibre optic light cable adapters are connected correctly.
2.
Connect the fibre optic light cables or check whether the fibre optic light cables are
connected properly.
3. Switch on the device and check whether all the LEDs on the ATMOS® LED Light Cube
CORIAN® control panel light up.
4.
Point each fibre optic light cable at a white surface and check whether the light
source lights up.