Patient Chair E 2

535.0000.B

535.0000.B with option: 

electrical backrest

2015-01  Index: 04

Operating Instructions

English

Содержание E 2

Страница 1: ...MedizinTechnik ATMOS Patient Chair E 2 535 0000 B 535 0000 B with option electrical backrest 2015 01 Index 04 Patient Chair E 2 Operating Instructions English ...

Страница 2: ...3 Power connection 8 4 0 Operating 9 12 4 1 Positioning the patient 9 4 1 Adjusting the seat height 9 4 2 Rotating the upper part of the chair 10 4 3 Rotating the seat 10 4 4 Adjusting the backrest 11 4 5 Back rest movement electrical 11 4 6 Adjusting the arm rests 11 4 7 Adjusting the head rest 12 4 8 Foot rest 12 4 9 Chassis optional 12 5 0 Cleaning 13 14 5 1 General information on cleaning and ...

Страница 3: ...art up please peruse chapter 2 0 For your safety in order to be prepared for any possible dangerous situations The patient chair has an electromotive height adjustment from 58 5 cm 78 5 cm The height adjustment is controlled by a foot switch The Auto Down homing function drives the chair automatically to its lowest position if so desired The upper part of the chair rotates and can be locked at any...

Страница 4: ...numeration Numeration General information Follow the arrows whilst proceeding sequence Checking Please read important information Important notes Symbols of the ATMOS patient chair E 2 Order number The CE sign shows that this product meets the appropriate requirements of the EC guidelines Application part type B REF Exchange Move plug in this direction Engage then check correct fit Turn shift in t...

Страница 5: ... and offering suffi cient protection Please document and report damages in transit immediately For complaints or return deliveries please use the enclosed form QD 434 The unit must be allowed to stand for up to six hours at room temperature prior to starting up for the fi rst time following transport at temperatures below freezing The unit may not be operated if it has not acclimatised as this mig...

Страница 6: ... with wet hands The ambient conditions specified in section 7 0 Technical specifi cations must be strictly observed The ATMOS patient chair E2 is not intended for operation in areas of medical rooms subject to danger of explosion Areas subject to explosion hazards may be created by the use of infl ammable anaesthetics skin cleansers and disin fectants Ensure that your patients sit in the middle of...

Страница 7: ...nt space is available for proper rotation of the upper part of the chair Before first operation of chair read and observe the safety instructions in section 2 0 Finally connect power cord 3 2 1 Front view Individually adjustable headrest Lever for infinitely variable synchronous adjustment of the backrest arm rests and foot support Arm rests to be swivelled off backwards Rotary seat Lever for arre...

Страница 8: ...power outlet features a proper ground connection Find all relevant electrical data voltage and nominal frequency as well as the data for fuses on the label located over the socket To disconnect the device pull the power cord out of outlet socket There is no indication that product is hooked up to line power Disconnect chair if not in use during service and repair work and or for cleaning 3 0 Insta...

Страница 9: ...riod of approx 8 5 min is re quired If continuous operation exceeds 1 5 min the thermo switch is activated and after that operation is only possible after approx 20 minutes cooling period 4 2 Adjusting the seat height Height adjustment of seat cushion level is controlled by 2 foot switches fig 2 Left foot switch up Right foot switch down Furthermore the ATMOS patient chair features an Auto Down ho...

Страница 10: ...ing angle is limited for chairs with an electrical back rest adjustment In this case the chair can be rotated by 150 either to the right or to the left 4 4 Rotating the seat The seat with the patient can be rotated independent of a backrest movement To do so adjust the lateral locking lever Fig 4 beyond its central position The seat can then be rotated 90 to the left or 90 to the right If the lock...

Страница 11: ...eir positions synchronously Fig 5 Lever element 4 6 Back rest movement electrical The arm rests foot rest and back rest are mechanically inter locked and move synchronized As an option electric actuators are available to move the back rest of the ATMOS E 2 patient chair The electric back rest movement is controlled by one foot switch Fig 6 in the base cover and two toggle switches located in the s...

Страница 12: ... 15mm Extend the chassis Move the chair upwards until the green light illumi nates fig 10 Push the lever to the left fig 12 to extent the chassis Move the chair downwards until the red light illumi nates fig 10 The chair can now be moved Retract the chassis Move the chair upwards until the green light illumi nates fig 10 Push the lever forward fig 11 to retract the chas sis Move the chair to the d...

Страница 13: ...s supply The unit itself can be wiped off with a moist not wet cloth Substances such as blood need to be removed immediately to prevent stains on the cover material of the uphol stery The described measures to clean and to disinfect or sterilize do not rule out regulations valid for the particular company In chapter 5 3 and 5 4 all surface and instrument disintectants are stated which are suitable...

Страница 14: ...ive ingredients QAV quarternary ammonium compounds 5 4 Recommended disinfectants for the upholstery Disinfectant Ingredients in 100 g Manufacturer ATMOS Green Clean SK Alkyldimethylbenzylammoniumchloride 1 g Metasys Rum Austria application solutionungslösung Dialkyldimehtylammoniumchloride 1 g Alkyldimethylethylbenzylammoniumchloride 1 g 5 0 Cleaning ...

Страница 15: ... to disconnect the device from line power 6 1 Fuse exchange To open the fuse socket rotate top counter clockwise Exchange of fuse To close fuse socket rotate top cw Malfunction Possible cause Solution Chair does not start up Mains plug is not connected correctly Check connection at the socket No mains voltage Check house fuse Check mains plug on perfect fit at the device Defective fuse Exchange th...

Страница 16: ...16 8 0 Accessories and Spare Parts REF 8 1 Accessories Child seat on request 8 2 Spare parts ...

Страница 17: ...ck rest inclination 7 bis 90 horizontal position Protective earth conductor resistance Earth leakage current Patient leakage current Enclosure leakage current n c 0 1 mA Ambient conditions Transport storage Operation 10 50 C 30 95 humidity without condensing air pressure 500 1060 hPa 10 35 C 30 95 humidity without condensing air pressure 500 1060 hPa Dimensions HxWxD 1490 mm x 630 mm x 930 mm Weig...

Страница 18: ...E and RoHS the device may not be disposed of in domestic waste Please observe existing national laws and rules for disposal of old devices Disposal within the Federal Republic of Germany In the Federal Republic of Germany the law for electrical devices ElektroG rules the disposal of electrical devices In order to guarantee a proper disposal of your old device please either pass on your old device ...

Страница 19: ...r 6 kV Contact 8 kV Air Floors should be wood concrete or ceramics tile If fl oors are synthetic the relative humidity should be at least 30 EFT IEC 61000 4 4 2 kV Mains 1 kV I Os 2 kV Mains 1 kV I Os Mains power quality should be that of a typical commercial or hospital environment Surges IEC 61000 4 5 1 kV symmetric 2 kV symmetric 1 kV symmetric 2 kV symmetric Mains power quality should be that ...

Страница 20: ...the test levels 11 3 Guidelines and Manufacturer s Declaration Immunity The ATMOS Patient chair E2 is intended for use in the electromagnetic environment specified below The customer or user of the ATMOS Patient chair E2 should ensure that it is used in such an environment Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance Conducted RF IEC 61000 4 6 3 Veff 150...

Страница 21: ... MHz the field strength is to be below 3 V m The ATMOS Patient chair E2 is intended for use in electromagnetic environment in which radiated disturbances are controlled The customer or user of the ATMOS Patient chair E2 can help prevent electromagnetic interfe rence by maintaining a minimum distance between portable and mobile RF Communications equipment and the ATMOS Patient chair E2 as recommend...

Страница 22: ...22 For your notes ...

Страница 23: ...23 For your notes ...

Страница 24: ...ds damage or loss is therefore transferred to the client as soon as the goods leave the factory or the client is in default of acceptance of the goods This also applies to cases where we confirm prepaid carriage Transport packaging and all other packaging according to the packaging regulations is not returnable Our client is responsible for disposing the packaging at its own cost Our deliveries ar...