1
1
1
1
?Cal,0
ASCII
?Cal,2
ASCII
?Cal,1
ASCII
?Cal,3
ASCII
Dec
Dec
Dec
Dec
0
Null
0
Null
0
Null
0
Null
1
Dec
0
Null
1
Dec
0
Null
1
Dec
0
Null
1
Dec
0
Null
300ms
processing delay
Wait 900ms
Wait 900ms
Wait 900ms
Wait 300ms
Wait 300ms
Calibration
or
or
Calibration
Issuing the cal,mid command after the
EZO
TM
pH circuit has been calibrated, will
clear the other calibration points. Full
calibration will have to be redone.
Command syntax
Cal,mid,7.00
Cal,low,4.00
Cal,high,10.00
Cal,clear
Example
Response
Cal,?
Cal,mid,n
Cal,low,n
Cal,high,n
Cal,clear
Cal,?
single point calibration at midpoint
two point calibration at lowpoint
three point calibration at highpoint
delete calibration data
device calibrated?
two point
one point
three point
50